The LR-Cal LCC 100 pressure controller measures and generates pressure and vacuum and consequently it is more than a mere measuring instrument - it can also be applied as test and calibration instrument for pressure sensors, pressure swithes and pressure gauges. Thanks to internal pressure/vacuum generation no auxiliary tools are required for operation. An internal Li-Ion battery (chargeable) makes mobile operation very easy.
Parallel operation of mains and battery supply allows for maximum flexibility.
Typical Applications:
Standard scope of delivery:
Type | LCC 100-1 | LCC 100-10 | LCC 100-100 | LCC 100-1000 | |
Order-Code | LCC-100-1 | LCC-100-10 | LCC-100-100 | LCC-100-1000 | |
Pressure range | -1 mbar | 10 mbar | 100 mbar | 1000 mbar | |
Overpressure | 5-fold | 5-fold | 5-fold | 2-fold | |
Measurement Uncertainty (+1 digit) | ±0.3% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS | |
Linearity ±1digit | ±0.2% FS | ±0.1% FS | ±0.1% FS | ±0.1% FS | |
Hysteresis | max. 0.1% FS | ||||
Measurement principle | inductive | ||||
Temperature drift of the internal reference sensor |
Zero: 0.003% FS / K (0% via zero balance) Span: 0.03% FS / K |
||||
Zero balance | • automatic (in settable time spans, possible to swith of), or • manuel (by pressing the ZERO button) |
||||
Long-term stability of the internal reference sensor |
0.1% FS per year (typical) | ||||
Temperature ranges | Working range +10...+40°C; Storage -10...+70°C | ||||
Usable pressure measuring range | -10...+110% | ||||
Adjustment accuracy | 0.01% FS | ||||
Adjustment time | depending on volume, <5 sec | ||||
Measurement media | Air, non-aggressive and corrosion-free dry gases | ||||
Operation modes | • CTRL (controlling pressure) • MESS (measuring pressure) • AUTO (individually definable pressure profile) • Remote-controlled operation (via USB or RS232 interface) |
||||
Display | Graphic display (white on dark blackground) | ||||
Interfaces | USB and RS232 | ||||
Supply | 24 VDC / 1 A, build-in Li-Ion battery (8h typical) and mains charger 90...264 VAC | ||||
Pressure ports | 6.6 x 11 mm (for flexible hose with 6 mm inner diameter) | ||||
Dimension | H 102,6 x W 257 x D 271 mm (without handel) |
The integrated electrical pump can make max. -600 mbar vacuum. If a larger vacuum is required, an external vacuum source is needed, e.g. the pressure test pump models LR-Cal LPP 08 and LR-Cal 2941.
The LR-Cal LCC 100 can make up to -10% of full range value as vacuum. If you have to measure/controll larger vacuum values, you can do this by exchaning the +/- ports.
Measure
The measuring mode ("MESS") is used to measure a gauge or differential pressure. In this operation mode the internal pump of the LR-Cal LCC 100 is deactivated. The applied pressure, measured by the internal reference sensor, is shown on the display.
Auto
In the "AUTO" mode, a calibration procedure can be defined. This mode allows a comfortable calibration of several test specimen with same specifications. The calibration can be run AUTOmatically in this mode.
Following graphic shows the principle of a calibration procedure:
(1) Wait for start command (pressing OK key)
(2) Delay time (adjustable)
(4) Duration of ZERO balancing of the system
(7) Dwell time at ZERO point
(9) Adjustment time to the next step
(10) Dwell time
(13) Delay time at max. pressure step
(15) Adjustment time to the next step
(16) Dwell time at ZERO point
(18) Delay time (adjustable)
Control
The control mode ("CTRL") is used to calibrate pressure transmitter, pressure switches or pressure gauges. In this operation mode the internal pump of the LR-Cal LCC 100 is activated, the pressure is adjusted to the set point. The pressure value is shown on the display.
Leak Test
In the CTRL mode, connected test specimen can be tested to leakage / tightness by pressing the LEAK key.
Order-Code | Description | Image |
LCC-100-KOFFER | Rigid carrying and storage case with custom foams | ![]() |
LCC-100-NT | SPARE PART: Mains adapter 90...264 VAC |