LR-Cal 2911 + 2941
Pressure test pumps
LR-Cal 2911: pressure up to +7 bar / LR-Cal 2941: Vacuum up to -0,9 bar
- Portable
- Pneumatic (with ambient air)
- LR-Cal 2911:
Test pressure generation up to 7 bar (100 psi), or - LR-Cal 2941:
Vacuum generation down to -0.9 bar (-25 inHg) - Simple generation, precise fine adjustment
Description
LR-Cal 2911 + LR-Cal 2941 description
The calibration hand test pumps LR-Cal 2911 is used for pressure generation and the LR-Cal 2941 is used for negative pressure/vacuum generation for checking, adjusting and calibrating mechanical and electronic pressure measuring instruments by comparison measurement. These tests can be carried out on a mobile basis, on site at the measuring point, or stationary in the laboratory or workshop.
If the measuring instrument to be tested and a sufficiently accurate reference measuring instrument are connected to the LR-Cal 2911 or LR-Cal 2941 calibration hand test pump via the T-piece supplied, the same pressure or vacuum acts on both measuring instruments when the pump is actuated. By comparing the two measured values at any pressure or vacuum values, it is possible to check the accuracy or adjust the vacuum gauge to be tested.
Despite their very compact dimensions, the LR-Cal 2911 and LR-Cal 2941 calibration hand test pumps enable simple and accurate pressure or vacuum generation. For precise adjustment for accurate comparison testing, the LR-Cal 2911 and LR-Cal 2941 have a fine adjustment valve. The reference pressure gauge and test item are connected via hose connection for hose with 4 mm internal diameter (6 mm external diameter). A piece of hose and a T-piece are included in the scope of delivery.
Specification
LR-Cal 2911 + LR-Cal 2941 specification
Pressure generation: | only type LR-Cal 2911: 0 to 7 bar (100 psi) |
Vacuum generation: | only typ LR-Cal 2941: 0 to -0,9 bar (-25 inHg) |
Pressure port: | For hose with 4 mm inner diameter and 6 mm outer diameter |
Medium: | Ambient air |
Pressure- resp. vacuum generation via: | Hand knob |
Test pressure fine adjustment: | Volume variator (very sensitive fine adjustment valve) |
Dimension: | approx. 200 mm long, pump body approx. 35 mm diameter |
Materials: | Aluminum, brass, H-PVC, NBR |
Weight: | approx. 400 g |
Order-Code: | LR-Cal 2911 (pressure generation): 500.9.000.1005.0 LR-Cal 2941 (vacuum generation): 500.9.000.9800.0 |
Scope of delivery
LR-Cal 2911 + LR-Cal 2941 scope of delivery
- Pressure test pump LR-Cal 2911 or LR-Cal 2941 with hose connection
- Test piece connection hose, length approx. 0.3 meter
- T-piece (for hose with inner diameter 4 mm and outer diameter 6 mm)
- Operating manual DE+EN
Note: For simultaneous connection of a device under test (DUT) and a reference instrument, two additional pieces of tubing are required, see accessories for suitable tubing by the meter.
LR-Cal 2911 + LR-Cal 2941 accessories + spare parts
Transport andstorage case
Accessory:
Rigid transport and storage case with molded foam insert.
Quick coupling T-piece for LR-Cal Serie 3000
Accessory:
T-piece with quick couplings
for hose with
1 x 6 mm outer diameter
2 x 4 mm outer diameter
Is needed if the pumps are used with digital handheld manometers LR-Cal Serie 3000.
PVC hose 4x6 mmper meter
Accessory:
PVC hose by the meter
Outer diameter 6 mm
(inner diameter 4 mm)
Hydrophobic filterfor hose 4x6 mm
Accessory:
Hydrophobic filter (max. 2 bar!)
gas permeable, blocks liquids/moisture,
for hose with
6 mm outer diameter
(4 mm inner diameter)
Quick coupling T-piece for hose 6 mm AD
Spare part:
T-piece with quick couplings
for hose with
6 mm outer diameter
(4 mm inner diameter)
Note: 1 pc. is included in scope of delivery of the pump.