DM 80-UMS
Digital pressure gauge with data logging
Accuracy ±0.25% BFSL, pressure ranges up to 2,500 bar
- Digital pressure gauge with measurement memory
- Accuracy ±0.25% BFSL
- -1...+16 to 0...2,500 bar
- USB interface
- Battery operated (chargeable)
Description
DM 80-UMS description
The Digital Manometer DM 80-UMS is very well suited for test and inspection tasks as well as for industrial applications.
The DM 80-UMS digital manometer is excellently suited for on-site use together with our LR-Cal pressure test pumps. In combination with our LR-Cal pressure comparison test pumps (spindle pumps), it is also very well suited for operation in the laboratory and in the workshop.
The device is equipped with a micro USB socket.
The measured value storage rate can be set as follows: every 10 ms / every 100 ms / every 1 sec / every 10 sec.
Furthermore, a data record can be created by pressing a button ("REC" key). Because of the exceedingly large storage capacity, the logging duration is as follows:
- with storage interval every 10 milliseconds: approx. 1 year
- with storage interval every 100 milliseconds: approx. 10.1 years
- with storage interval every 1 second: approx. 101.5 years
- with storage interval every 10 seconds: approx. 1,015 years
The built-in data memory of the DM 80-UMS is recognized by the PC as external memory, like an USB stick. This allows extremely easy readout of the stored data (.csv files) without special software. The data transfer speed corresponds to the USB 2.0 standard.
The built-in rechargeable lithium-ion battery allows a self-sufficient continuous operation of about 10 hours. The charging time of an empty battery is about 3 hours.
Specification
DM 80-UMS Spezifikation
Accuracy | ±0.25% BFSL *) |
Interface | USB |
Data memory | 8 GB (CSV files) |
Ambient temperature | -10...+50°C |
Storage temperature | -20...+60°C |
Medium temperature | -40...+125°C |
Max. relative humidity | 85% r.h. |
Material pressure connection | Stainless steel 1.4301 |
Material membrane | Stainless steel 1.4548 (welded) |
Material case | Metal (die-cast zinc) |
Temperature compensated range | 0...50°C |
Error in temperature compensated range | ≤0.035% FS / °C |
Minimum load cycles | 100 x 106 |
Supply | Lithium-Ion battery |
Max. battery life time | 10 h continuous operation |
Pressure units | switchable: bar, PSI, kPa, MPa (depending from measuring range) |
Pressure connection | 1/2" BSP male according to EN 837 |
Display | LCD incl. bargraph |
Peak values | Min + Max |
Weight | approx. 610 g |
Case diameter without rubber protection cap | approx. 80 mm |
Case diameter with obligatory rubber protection cap | approx. 95-100 mm |
*) ±0.25% BFSL (maximum nonlinearity according to best fit straight line method) corresponds to ±0.5% FS according to IEC 60770 (limit point setting (nonlinearity, hysteresis and repeatability)).
Pressure ranges
DM 80-UMS pressure range
Order-Code | Measuring range [bar] | Pressure type |
DM80-UMS-0025-0-G12 | 0...25 | gauge |
DM80-UMS-0040-0-G12 | 0...40 | gauge |
DM80-UMS-0060-0-G12 | 0...60 | gauge |
DM80-UMS-0100-0-G12 | 0...100 | gauge |
DM80-UMS-0250-0-G12 | 0...250 | gauge |
DM80-UMS-0400-0-G12 | 0...400 | gauge |
DM80-UMS-0600-0-G12 | 0...600 | gauge |
DM80-UMS-1000-0-G12 | 0...1,000 | gauge |
DM80-UMS-2500-0-G12 | 0...2,500 | gauge |
DM80-UMS-9016-0-G12 | -1...+16 | gauge |
Absolute pressure measuring ranges are not available for type DM 80-UMS.
Scope of standard supply
DM 80-UMS scope of standard supply
- Digital pressure gauge with data memory type DM 80-UMS
- Built-in Lithium-Ion battery, chargeable
- Built-in 8 GB data memory
- USB cable (Micro-USB to USB-A)
- Factory certificate of calibration
- Operating manual DE oder EN
1 Rubber protection cap code "DM80-UMS-GSK-G" must be ordered with each unit!
DM 80-UMS accessories
Rubber protection cap
Accessory:
Green rubber protection cap
1 piece must be ordered with each DM 80-UMS
Rubber protection cap with hook
Accessory:
Green rubber protection cap
with hook for hanging
and wire rope 340 mm long
Thread adapter M16x2 F x 1/2" BSP F
Accessory:
Thread adapter
1/2" BSP female x M 16 x 2 female
Thread adapter M16x2 M / G 1/2 F
Accessory:
Thread adapter
1/2" BSP female x M 16 x 2 male
Carrying case
Accessory:
Storage and transport case with molded foam insert
Charger110...230 VAC
Accessory:
Plug-in charger 110...230 VAC