Dry well Temperature calibrator
Model LR-Cal SOLAR-1200
Range +200°C...+1200°C
The portable dry well (metal block) temperature calibrator LR-Cal SOLAR-1200 is temperature source and reference instrument in one device, for checking and calibrating of all kind of temperature instruments, such as thermometer, RTDs, Thermocouples, Temperature switches, etc.
- Temperature range +200...+1200°C
- Ramp function
- Thermostat tests
- Computer aided calibration via computer interface
- Several ceramic / Inconel inserts available with different holes
The ceramic equalizing block, whose large size suits the external dimensions, ensures a proper heat transmission as well as optimal stability and uniformity values. It's interchangeability makes it possible to test sensors of various lengths, with diameters ranging from 1 mm up to 26 mm.
The device is equipped with a counter-current forced air cooling system, which keeps the temperature in the upper part of the well low. This feature allows checking also of very short probes, without heating the connecting head or the handgrip.
Specification
LR-Cal SOLAR-1200 Specification
Temperature range |
+200°C...+1200°C |
Temperature indication |
Accuracy ±3°C Resolution 0,1° (°C or °F or K) |
Stability of regulated temperature |
±0.3°C (at 1000°C) |
Heating |
max. 15°C per minute |
Cooling |
max. 7°C per minute |
Temperature uniformity |
at 1000°C: radial ±2°C axial ±03°C |
Block depth |
160 mm |
Block diameter |
35 mm (block made in ceramic) |
Power supply |
230 VAC |
Power consumption |
700 VA |
Interface |
RS232 (optional RS232-USB converter available) |
Housing material |
Metal |
Weight |
approx. 10.5 kg |
Dimension |
350 x 160 x 390 mm |
Scope of delivery + Accessories
LR-Cal SOLAR-1200 Scope of delivery + Accessories
Scope of standard delivery:
- Portable dry well temperature calibrator LR-Cal SOLAR-1200
- Spare fuse
- Power cable
- 1 ceramic block with 4 holes: 4.8 - 6.4 - 9.5 - 12.8 mm diam.
- Upper insulation with 4 holes, suitable to supplied ceramic block
- Twizzers for removing insert
- Operating manual
- Certificate / Test Report (accuracy and stability)
Optional Accessories:
Order-Code |
Description |
Image |
REF-T-SENSOR-S |
External reference thermocouple type S Temperature range 0...1300°C |
REF-T-VGLST-S |
Compensation in metal tube of external reference probe |
DKD-T-KAL-S |
DAkkS certificate of calibration for external reference probe |
KONST-T-KAL-S |
Calculation of constants for external reference probe |
KONST2-T-KAL-S |
alternatively: calculation of constants incl. table with values (°C steps) |
511.0.999.0008.0 |
Rigid case, aluminium 400 x 300 x 500 mm |
 |
599.0.000.0003.0 |
PC software LR-Cal AQ2sp incl. RS232-interface lead Temperatur sensor calibration: Automatic calibration routine for all type of temperature sensors with advanced features such reference sensor stabilization, time-calibration data are automatically populated and stored for printing report. This procedure can also be ran in manual mode to allow the operator to manually enter the data. This is particular useful when calibrating temperature device such as digital thermometer or analogue thermometers. Stress test: Simply set MAX and MIN value and the software will run an automatic stress test on single sensor or a batch. Advance thermostat calibration: Simply set the open and close temperature limits, type of thermostat, typ of test and the software will automatically run the test. |
 |
RS232-USB-KONV |
RS232 - USB interface converter |
|