LR-Cal QUARTZ-50
Dry block temperature calibrator
-50...+150°C
Description
LR-Cal QUARTZ-50 description
The LR-Cal QUARTZ-50 dry block temperature calibrator serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments.
- Temperature range -50...+150°C
- Ramp function
- Thermostat tests
- Computer-controlled calibrations via computer interface (PC software optionally available)
- Various dry block inserts available with different holes for test specimens
- Dry block insert "Black Body" for testing and calibration of non-contact infrared thermometers
- Incl. factory certificate (accuracy, stability, performance)
- Compunter interface (RS232, interface converter to USB available as accessory)
Device versions
LR-Cal QUARTZ-50 device versions
The dry block temperature calibrator LR-Cal QUARTZ-50 can be supplied in two different versions:
Version LR-Cal QUARTZ-50
Standard version
Order-Code QUARTZ-50
Without inputs for resistance thermometer / thermocouples
Version LR-Cal QUARTZ-50-2I
Version with 2 measuring inputs for resistance thermometers Pt 100 (2-, 3- or 4-wire) or Pt 1000 as well as thermocouples incl. cold junction compensation types B, E, J, K, N, R, S and T.
Order-Code QUARTZ-50-2I
The signals from up to two external temperature sensors can also be displayed.
Extended scope of delivery:
Electrical connection cables, plugs/terminals for the electrical connections.
Accuracy (maximum deviation) of the two measuring inputs for device version "-2I": | ||||||||
Pt 100 or Pt 1000 | at -40°C: | ±0.09°C | at 0°C: | ±0.08°C | at +150°C: | ±0.11°C | at +300°C: | ±0.14°C |
Thermocouple type B | at +950°C: | ±0.97°C | at +1050°C: | ±1.03°C | at +1200°C: | ±1.12°C | ||
Thermocouple type E | at -40°C: | ±0.42°C | at 0°C: | ±0.40°C | at +350°C: | ±0.61°C | ||
Thermocouple type J | at +200°C: | ±0.52°C | at +450°C: | ±0.67°C | at +700°C: | ±0.82°C | ||
Thermocouples types K, N, R and S | at +400°C: | ±0.64°C | at +700°C: | ±0.82°C | at +1000°C: | ±1.00°C | ||
Thermocouple type T | at -40°C: | ±0.42°C | at 0°C: | ±0.40°C | at +350°C: | ±0.61°C |
Specification
LR-Cal QUARTZ-50 specification
Temperature range (at +20°C ambient temperature) | -50°C...+150°C | |
Accuracy of temperature indication | ±0.15°C | |
Resolution of temperature indication | 0.01° / 0,1° | |
Temperature units (selectable) | °C / °F / K | |
Stability of regulated temperature | ±0.03°C at 0°C | |
Mean heating time | 3.5°C / min. | |
Mean cooling time | 1°C / min. | |
Temperature uniformity | radial: ±0.02°C (at 0°C at 40 mm from the bottom) axial: ±0.1°C (at 0°C and 40 mm from bottom of the holes of the insert) |
|
Insertion depth | 135 mm (holes depth of inserts up to 125 mm) | |
Usable block diameter | 35 mm | |
Power supply | 230 VAC (50/60 Hz) | |
Power consumption | 400 VA | |
Interface | RS232 (USB converter available as option) | |
Housing material | Metal | |
Weight | approx. 9.3 kgs | |
Dimension |
350 x 160 x 360 mm |
|
Order-Code (standard version) | QUARTZ-50 | |
Order-Code (-2I version with 2 measuring inputs) | QUARTZ-50-2I |
Scope of standard supply
LR-Cal QUARTZ-50 scope of standard supply
- Portable dry block temperature calibrator LR-Cal QUARTZ-50 or LR-Cal QUARTZ-50-2I
- Spare fuses
- Cable for power supply
- Connection cable for thermostat tests
- 1 dry block insert with 6 holes (3.5 + 4.5 + 5.5 + 6.5 + 8.5 + 10.5 mm diameter)
- Removing tool for changing dry block inserts
- Factory certificate (accurace, stability, performance)
- Operating manual EN
Additionally for version LR-Cal QUARTZ-50-2I:
- Electrical connection cables (red/black) and screw terminals (red/black) for the two measurement inputs
Available dry block inserts
LR-Cal QUARTZ-50 available dry block inserts
Order-Code | Description |
507.0.999.0008.0 | Blank dry block insert without holes (for self-drilling) |
507.0.999.0009.0 | Dry block insert with 6 holes: 3.5 + 4.5 + 5.5 + 6.5 + 8.5 + 10.5 mm diameter 1 piece already included in the device scope of delivery |
507.0.999.0010.0 | Dry block insert with 2 holes: 6.5 + 19.5 mm diameter |
INS-SPEC-633-3 | Dry block insert with 1 to 3 holes according to customer requirements (3.5 to 20 mm diameter, if feasible) |
INS-SPEC-633-6 | Dry block insert with 4 to 6 holes according to customer requirements (3.5 to 20 mm diameter, if feasible) |
INS-SPEC-633-10 | Dry block insert with 7 to 10 holes according to customer requirements (3.5 to 20 mm diameter, if feasible) |
INSERT-QUARTZ | Dry block insert "Black Body", 26 mm inner diameter, for infrared thermometer Note: For instrument version "Standard", reference thermometer LR-Cal LRT 750 is additionally required For instrument version "-2I" additionally Pt 100 reference sensor cl. A, code BB-SENSOR-01 is required. |
LR-Cal QUARTZ-50 other accessories + spare parts
DAkkS certification
Option:
DAkkS certification for the measuring chain consisting of display of the calibrator and internal reference sensor at 6 measuring points
Shoulder bag
Accessory:
Padded shoulder bag
for LR-Cal QUARTZ-50(-2I)
Dimensions: 410 x 220 x 380 mm
Weight: 1.45 kg
Shipping dimensions: 600 x 370 x 500 mm
Case
Accessory:
Case with molded foam inserts
for LR-Cal QUARTZ-50(-2I)
Dimensions: 450 x 280 x 380 mm
Weight: 4.8 kg
Shipping dimensions: 600 x 370 x 500 mm
Windows-PC software
Accessory:
Windows PC software AQ2Sp2
Download version
incl. RS232 connection cable
RS232-USB Converter
Accessory:
Interface converter
RS232 to USB
Insert removing tool
Spare part:
Tool for removing dry block inserts
(1 piece already included in the scope of delivery of the device)
The test item (temperature measuring instruments of all types, e.g. dial thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.
Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.
The LR-Cal metal dry block temperature calibrators have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.
Furthermore, the LR-Cal AQ2sp2 PC software is available as an accessory for the LR-Cal metal block temperature calibrators. Herewith the calibrator can be controlled completely from the PC (if necessary with optional RS232/USB converter) via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.
Synonyms
Metal dry block temperature calibrators are also referred to as
- Dry-well calibrator
- Temperature dry-well calibrator
- Dry block calibrator
- Temperature calibrator
- Temperature calibration device
- Temperature calibration oven