Skip to main content

LR-Cal PULSAR-35Cu
Dry block temperature calibrator
Ambient temperature ...+600°C

LR-Cal PULSAR-35Cu dry well temperature calibrator
LR-Cal
  • Portable dry block temperature calibrator
  • Range: ambient temperature to +600°C
  • Accuracy ±0.3°C, stability ±0.05°C
  • Block: 35 mm diameter
    Insertion depth: 185 mm
Dry block inserts for LR-Cal PULSAR-35Cu

Dry block inserts for
LR-Cal PULSAR-35Cu

Ansicht von oben LR-Cal PULSAR-35Cu

Ansicht von oben
LR-Cal PULSAR-35Cu

LR-Cal PULSAR-35Cu other accessories + spare parts

DAkkS certification

Code: DKD-T-KAL-TK-6T
DKD-T-KAL-TK-6T

Option:
DAkkS certification for the measuring chain consisting of display of the calibrator and internal reference sensor at 6 measuring points

Shoulder bag

Code: TEMPKAL-TASCHE
TEMPKAL-TASCHE

Accessory:
Padded shoulder bag
for LR-Cal PULSAR-35Cu(-2I)
Dimensions: 410 x 220 x 380 mm
Weight: 1.45 kg
Shipping dimensions: 600 x 370 x 500 mm

Case

Code: TEMPKAL-KOFFER
TEMPKAL-KOFFER

Accessory:
Case with molded foam inserts
for LR-Cal PULSAR-35Cu(-2I)
Dimensions: 450 x 280 x 380 mm
Weight: 4.8 kg
Shipping dimensions: 600 x 370 x 500 mm

Windows-PC software

Code: 599.0.000.0003.0
599.0.000.0003.0

Accessory:
Windows-PC software AQ2Sp2
Download version
incl. RS232 connection cable

RS232-USB converter

Code: RS232-USB-KONV
RS232-USB-KONV

Accessory:
Interface converter
RS232 to USB

Insert removing tool

Code: 599.0.000.0001.0
599.0.000.0001.0

Spare part:
Tool for removing dry block inserts

(1 piece already included in the scope of delivery of the device)

Explanation of dry block temperature calibrators

The test item (temperature measuring instruments of all types, e.g. dial thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.

Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.

The LR-Cal metal dry block temperature calibrators have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.

Furthermore, the LR-Cal AQ2sp2 PC software is available as an accessory for the LR-Cal metal block temperature calibrators. Herewith the calibrator can be controlled completely from the PC (if necessary with optional RS232/USB converter) via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.

Synonyms

Metal dry block temperature calibrators are also referred to as

  • Dry-well calibrator
  • Temperature dry-well calibrator
  • Dry block calibrator
  • Temperature calibrator
  • Temperature calibration device
  • Temperature calibration oven