Accessibility Tools

Skip to main content

Terminology Calibration

Terminology "Calibration"

Calibration

Calibration is the determination and documentation of deviations in a measuring device. This measuring device is compared with the measurement result of another measuring device (or standard), whereby the standard is traceable to national or international standards and therefore its correct and accurate function is known.

Adjustment

Adjustment involves intervening in the device to minimize a detected measurement deviation. Normally, a further calibration is carried out after an adjustment to verify that the intervention in the device was correct.

Traceability

Traceability involves comparing the measured value of a measuring device with a national or international standard. At each step, according to the hierarchy of standards, a measuring device is compared with a standard, which in turn has been calibrated with a higher-level standard. The calibration hierarchy ranges from company test equipment, working or factory standards, to reference standards and national standards.

Measurement uncertainty

Measurement uncertainty is a non-negative parameter and is specified without a sign. It characterizes the dispersion of the values associated with the measured quantity based on the information used.

Read more …Terminology Calibration

  • Hits: 331

LR-Cal LTC-FB-9035-45 Temperature micro bath -35...+120°C

LR-Cal LTC-FB-9035-45
Temperature calibration micro bath -35...+120°C

Brochure

LR-Cal LTC-FB-9035-45 temperature calibration micro bath LR-Cal
  • Portable
  • Temperature range -35°C...+120°C
  • Accuracy ±0.15°C, stability ±0.02°C
  • Bath volume approx. 300 cm³
    (45 mm diameter, 170 mm depth)
  • Option: changeable calibration bath insert
  • Temperature calibration micro bath LR-Cal LTC-FB-9035-45 with support for test item

    FLUID

    incl. support for test item and reference

  • Temperature calibration microbath LR-Cal LTC-FB-9035-45 touchscreen

    FLUID

    Simple operation via touchscreen

  • Exchangeable calibration bath insert for LR-Cal LTC-FB-9035-45

    FLUID

    LTC-FB-INSERT
    Changeable calibration bath insert

  • LTC-FB-ENT-TOOL

    FLUID

    LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL
    Optional tool set for removing protective grilles + magnetic mixer rod (easier emptying+cleaning)

  • Description

    LR-Cal LTC-FB-9035-45 Description

    The LR-Cal LTC-FB-9035-45 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

    • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
    • Homogeneous temperature distribution due to adjustable magnetic mixer
    • Simple operation with color touchscreen
    • Ramp function + thermostat tests
    • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
    • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
    • Option: dry block inserts to convert the device to a dry block temperature calibrator
  • Options LTC-MP-3I (3 measuring inputs) + LTC-MP-115V (for 115 VAC)

    Option LTC-MP-3I - three configurable measuring inputs

    Option LTC-MP-3I: 3 konfigurierbare MesseingängeThe LTC-MP-3I option can be ordered together with the LR-Cal LTC-FB-9035-45:

    Device version with 3 configurable measuring inputs. Of these, 2 for resistance thermometers (Pt 100 (2-, 3- or 4-wire) or Pt 100, or thermocouples incl. cold junction compensation, types  B, E, J, K, N, R, S and T, and 1 measuring input for analog output signals 4...20 mA or 0...10 V.
    Order-Code: LTC-MP-3I (must be ordered together with the base unit)
    Extended scope of delivery:
    Electrical connection cables, plugs/terminals (red/black) for electrical connections

    Accuracy (max. deviation) of resistance and thermocouple measurement inputs with option LTC-MP-3I:
    Pt 100 or Pt 1000 at -40°C: ±0.09°C at 0°C: ±0.08°C at +150°C: ±0.11°C at +300°C: ±0.14°C
    Thermocouple type B at +950°C: ±0.97°C at +1050°C: ±1.03°C at +1200°C: ±1.12°C
    Thermocouple type E at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C
    Thermocouple type J at +200°C: ±0.52°C at +450°C: ±0.67°C at +700°C: ±0.82°C
    Thermocouple types K, N, R and S at +400°C: ±0.64°C at +700°C: ±0.82°C at +1000°C: ±1.00°C
    Thermocouple type T at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C

    Option LTC-MP-115V - device version for für 115 VAC

    The LTC-MP-115V option can be ordered together with the LR-Cal LTC-FB-9035-45:

    Device version for power supply 115 VAC 50/60 Hz (instead of für 230 VAC).

  • Specification

    LR-Cal LTC-FB-9035-45 specification

    Description: Specification:
    Temperature range: -35...+120°C
    Display + operation: color touchscreen
    Display accuracy: ±0.15°C *
    Measuring units: °C / °F / K
    Displey resolution: 0.1° / 0.01° / 0.00^°
    Mean heating time:
    (incl. stabilization)
    approx. 40 min.
    from ambient temperature to +120°C
    Mean cooling time:
    (incl. stabilization)
    approx. 100 min.
    from +120°C to ambient temperature
    Stability: ±0.02°C at 0°C **
    Axial uniformity: ±0.05°C
    Radial uniformity: ±0.02°C at 0°C ***
    Hole diameter: 45 mm
    Hole depth: 170 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4.5 VDC
    Adjustable ramp function: from 0.1°C/min
    Computer interface: USB type B
    Automatic calibration: at 5 temperature points
    External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
    Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric powert: 350 VA
    Dimensions: 160 x 360 x 350 mm
    Weight incl. standard accessories: 11.0 kg

    All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
    The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 40 mm.

  • Scope of standard delivery

    LR-Cal LTC-FB-9035-45 scope of delivery

    • Portable temperature calibration micro bath LR-Cal LTC-FB-9035-45
    • Spare fuses
    • Power supply cable
    • Connection cables for thermostat tests
    • Bottle with 500 cm³ silicone oil 47V10
    • Holding support for test items, e.g. glass thermometers
    • Sealing cap for the reservoir (for transport purposes)
    • Carrying bag with shoulder trap
    • Only if option LTC-MP-3I is ordered with the device: electrical connection cables and plugs (red/black) for the three measuring inputs
    • Acceptance test certficate 3.1 in accordance with DIN EN 10204 (factory calibration certificate)
    • Operating manual DE+EN
  • Options + Accessories

LR-Cal LTC-FB-9035-45 options + accessories

  • LTC-MP-3I

    3 configurable
    measuring inputs

    Code: LTC-MP-3I

    Option, must be ordered together with the calibrator:
    Device version with 3 configurable measuring inputs
    2 x for Pt 100 / Pt 1000 / Thermocouples
    1 x for analog signal 4-20mA / 0-10V

  • LTC-MP-115V

    Power supply
    115 VAC

    Code: LTC-MP-115V

    Option, must be ordered together with the calibrator:
    Device version for power supply
    115 VAC 50/60 Hz
    (instead of 230 VAC)
    NO extra charge

  • LTC-FB-INSERT

    Exchangeable
    calibration bath insert

    Code: LTC-FB-INSERT

    Option:
    Calibration bath insert incl. closing cap
    For fast change of the silicone oil resp. the calibration fluid (without empying and cleaning the reservoir)

  • LTC-FB-COOL


    Cooling coil

    Code: LTC-FB-COOL

    Accessory:
    Stainless steel cooling coil. The cooling coil shortens the cooling time, and it can also shift the start of the working range of the calibrator downwards (but above 0°C).
    Requires fresh water connection

  • LTC-AQ2SP2

    Windows-PC
    software Aq2Sp2

    Code: LTC-AQ2SP2

    Accessory:
    Windows-PC software Aq2Sp2
    (delivered as download link)

    Download Manual (PDF)

  • LTC-FB-DB-45-0

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-45-0

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 44.7 mm
    Depth 170 mm
    Without holes (for self-drilling)

  • LTC-FB-DB-45-7

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-45-7

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 44.7 mm
    Depth 170 mm
    With 7 drillings/holes:
    4.0+4.5+5.5+6.5+8.5+10.5 mm

  • LTC-FB-DB-45-9

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-45-9

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 44.7 mm
    Depth 170 mm
    With 9 drillings/holes: 3.5+4.5+4.5
    +5.5+6.5+6.5+6.5+8.5+10.5 mm

  • LTC-FB-DB-KOFFER

    Sturdy case
    Aluminum

    Code: LTC-FB-DB-KOFFER

    Accessory:
    Sturdy case made of aluminum with molded foam inserts
    Dimension 450 x 280 x 380 mm
    Weight 4.8 kg
    Shipping dimensions:
    600 x 370 x 500 mm

  • LTC-FB-DB-TASCHE

    Carrying bag
    with shoulder trap

    Code: LTC-FB-DB-TASCHE

    Spare part:
    Carrying bag with shoulder trap
    Dimension 410 x 220 x 380 mm
    Weight 1.45 kg
    Shipping dimensions:
    600 x 370 x 500 mm
    (1 bag ist included in calibrators delivery)

  • LTC-FB-ENT-TOOL

    Removal tools for
    protect. grill + mixer rod

    Code: LTC-FB-ENT-TOOL

    Accessory:
    Removal tools for protective grilles and magnetic mixer rod
    Facilitates emptying and cleaning
    (Already included in the scope of delivery of LTC-FB-INSERT and LTC-FB-DB-... inserts)

  • DKD-T-KAL-TK-6T

    DAkkS-
    certification

    Code: DKD-T-KAL-TK-6T

    Option:
    DAkkS certificate of calibration

    For the measuring chain consisting of display of the calibrator and reference sensor, 6 measuring points

  • LTC-FB-SO-005

    Silikone oil
    47V5 (200C5)

    Art.Nr.: LTC-FB-SO-005

    Accessory:
    500 cm³ calibration fluid
    Silicone oil 47V5 (200C5)
    for temperature -40...+130°C

  • LTC-FB-SO-010

    Silikone oil
    47V10

    Code: LTC-FB-SO-010

    Spare part:
    500 cm³ calibration fluid
    Silikone oil 47V10
    for temperature -30...+150°C
    (1 bottle included in calibrators scope of delivery)


Suitable reference sensors (probes) for calibrator version with option LTC-MP-3I (3 configurable measuring inputs):

  • What is temperature calibration with a temperature calibration bath?

    What is temperature calibration with a temperature calibration bath?

    The test item (temperature measuring instruments of all types, e.g. dial thermometers, glass thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.

    Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.

    The LR-Cal temperature calibration baths have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.

    The LTC-MP-3I option is available for the portable LR-Cal LTC temperature calibration baths (with the exception of the LR-Cal LTC 150/250-F(-PLUS) devices), with three measuring inputs (2 x for RTD/TC and 1 x for analog signal 4...20 mA/0...10 V).

    The stationary LR-Cal temperature calibration baths are available in two versions: standard version and “-2I” version with 2 measuring inputs (for RTD/TC).

    Furthermore, the LR-Cal Aq2Sp2 PC software is available as an option for the LR-Cal temperature calibration baths. With this, the calibrator can be completely controlled from the PC via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.

    Synonyms

    Temperature calibration baths are also referred to as

    • Calibration bath
    • Micro bath
    • Temperature bath
    • Temperature calibrator
    • Temperature calibration device
    • Temperature calibration bath
  • Information about operating fluids for temperature calibration baths

    Calibration liquids for LR-Cal temperature calibration baths

    Applications:

    • Thermostatic fluids for micro baths and thermostatic temperature calibration baths

    Main Characteristics:

    • Operating range -50...+200°C
    • High stability
    • Good thermal conductivity
    • Nearly odourless
    • Long service live
    • Packaged in bottle or canister
    • Different types for a wide range applications
    • Low toxicity
    • Low viscosity and corrosovity

    Our thermostatic fluids are suitable for use in liquid thermostatic calibrators and thermostatic baths for applications from -50°C to +200°C.
    The choice of thermostatic fluid is the most important aspect for obtaining the best results with regard to operation and temperature control of calibrators and baths.
    These fluids have perfect thermodynamic characteristics to guarantee the best performance in the various fields of application, ensuring safe and reliable operation as well as optimum heat transfer efficiency.
    Silicone-based fluids are chemically inert substances that do not affect metals, are resistant to weathering and, when used correctly, the very low formation of cracking and oxidation ensures a long service life.

    Most of our thermostatic fluids are available in different containers: 500 cm³ bottles or 9 kg canisters (25 kg on request).

    Temperature application ranges of the various operating fluids:

    Type of calibration
    medium (liquid)
    Usable
    range
    (1)
    Temperature above which fume hood
    should be provided
    (2)
    Fume
    burning
    (3)
    A viscosity of 10 cst exists at a temperature of
    (4)
    Freezing temperature
    (5)
    Glycol 56% + water * -40...+90°C +70°C >110°C  +20°C -43°C
    Silicone oil 47V5 -40...+130°C +110°C +136°C  +5°C  -65°C
    Silicone oil 47V10 -30...+150°C +120°C +160°C  +20°C  -65°C
    Silicone oil 47V20 -20...+200°C ** +130°C +230°C  +60°C  -60°C
    Silicone oil 47V50 +30...+250°C ** +140°C +280°C  +160°C  -55°C
    Silicone oil 47V100 +50...+200°C ** +170°C >300°C  +200°C  -55°C
    Silicone oil 47V150 +50...+200°C ** +190°C >300°C  -/-  -55°C
    Note: Note text:
    * Mixture of 56% Glycol + 44% water
    ** In order to avoid gelling, it is recommended not to exceed a temperature of 200°C for a long time; even for fluids with wider operating ranges
    (1) Usable range
    The “usable range” of a calibration liquid is the temperature range in which it can be used under optimum conditions. The range may be limited by viscosity, flash points, freezing points, boiling points, evaporation rates, tendency to gel (or polymerize), etc.
    There is no liquid that covers extremely wide temperature ranges. Therefore, you should ideally use a separate calibration liquid for each temperature range. The recommended range is the one in which the viscosity is not optimal, but still reaches values that allow proper use.
    (2) Use of a fume hood
    The use of an exhaust hood when using temperature calibration baths prevents users from inhaling vapors from calibration liquids. It is best to use extraction devices near the access opening to the calibration bath. Vapors can otherwise settle on the mucous membranes of the eyes or on the skin, causing discomfort. Silicone oils can form benzene and formaldehyde when they decompose at high temperatures. The table shows the temperature above which an extractor hood should be used.
    (3) Flash point
    The table shows the temperature value at which the presence of an open flame can ignite vapours combustion.
    (4) Viscosity
    Viscosity is a measure of the flow resistance of a liquid. Viscosity increases with decreasing temperature. Kinetic viscosity is the ratio of absolute viscosity to density. This is measured in stokes and centistokes (cSt). 1 cSt = 1 stoke divided by 100. The higher the cSt, the more viscous or dense the liquid. Excessively viscous liquids place a strain on the stirring and pumping mechanisms of the calibration bath and do not transfer heat sufficiently. The table shows the temperature value below which the performance of the temperature calibration bath is no longer optimal. The equivalent value of water at 20°C is 1.1 cst.
    (5) Freezing point
    Silicone oils also change their state below a certain temperature by solidifying. The table shows the value at which the calibration fluid changes its state and solidifies.

    Other physical and chemical properties

    Specific heat
    The specific heat is the amount of heat required to raise or lower the temperature by 1 °C per kg. The specific heat, called CP, is expressed in Kcal/Kg °C. The energy required to heat or cool the calibration liquids depends on the amount of liquid contained in the calibrator in kg and its specific heat. A high specific heat value requires more energy to change the temperature, but offers greater temperature stability. The average CP of silicone fluids is 35 Kcal/Kg °C. The CP of water at 20 °C is equal to 1.

    Thermal conductivity
    Thermal conductivity is the ability of the liquid to transfer heat from one molecule to another. The better the heat transfer, the faster the liquid heats or cools. Better thermal conductivity improves the temperature uniformity in the reservoir of the calibration bath.

    Expansion
    All liquids have a coefficient of thermal expansion. This indicates how much the volume of a liquid changes (expands or contracts) when the temperature changes. Liquid expansion has an important impact on the safety, cleanliness and maintenance of temperature calibration baths. If calibrators are filled with too much liquid at a low temperature and then heated without regard to the increase in volume, they can overflow.

    If too little liquid is filled in the reservoir, below the level of the heating elements, the temperature rise of the non-immersed part of the heating elements could ignite the liquid.

    Gelation (polymerization)
    The gelling temperature is the temperature at which silicone fluids oxidize, gel and turn into a viscous slurry. The main cause of this phenomenon is oxidation. Although silicone fluids can be used safely up to their flash point, the susceptibility to polymerization increases when used above their oxidation point. To prevent polymerization, it is advisable to keep the operating time above the oxidation point of the silicone oil used as short as possible in order to keep contaminants (salts, other oils and oxidizing agents) away. The calibration fluid must be replaced if it becomes dark, viscous or temperature unstable. It is important to limit the application time at high temperatures to prevent the oil from degrading and to extend its service life.

Read more …LR-Cal LTC-FB-9035-45 Temperature micro bath -35...+120°C

  • Hits: 1052

LR-Cal LTC-FB-9018-60 Temperature micro bath -18...+140°C

LR-Cal LTC-FB-9018-60
Temperature calibration micro bath -18...+140°C

Brochure

LR-Cal LTC-FB-9018-60 temperature calibration micro bath LR-Cal
  • Portable
  • Temperature range -18°C...+140°C
  • Accuracy ±0.15°C, stability ±0.02°C
  • Bath volume approx. 500 cm³
    (60 mm diameter, 170 mm depth)
  • Option: changeable calibration bath insert
  • Temperature calibration micro bath LR-Cal LTC-FB-9018-60 with support for test item

    FLUID

    incl. support for test item and reference

  • Temperature calibration microbath LR-Cal LTC-FB-9018-60 touchscreen

    FLUID

    Simple operation via touchscreen

  • Exchangeable calibration bath insert for LR-Cal LTC-FB-9018-60

    FLUID

    LTC-FB-INSERT
    Changeable calibration bath insert

  • LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL
    Optional tool set for removing protective grilles + magnetic mixer rod (easier emptying+cleaning)

  • Description

    LR-Cal LTC-FB-9018-60 Description

    The LR-Cal LTC-FB-9018-60 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

    • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
    • Homogeneous temperature distribution due to adjustable magnetic mixer
    • Simple operation with color touchscreen
    • Ramp function + thermostat tests
    • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
    • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
    • Option: dry block inserts to convert the device to a dry block temperature calibrator
  • Options LTC-MP-3I (3 measuring inputs) + LTC-MP-115V (for 115 VAC)

    Option LTC-MP-3I - three configurable measuring inputs

    Option LTC-MP-3I: 3 konfigurierbare MesseingängeThe LTC-MP-3I option can be ordered together with the LR-Cal LTC-FB-9018-60:

    Device version with 3 configurable measuring inputs. Of these, 2 for resistance thermometers (Pt 100 (2-, 3- or 4-wire) or Pt 100, or thermocouples incl. cold junction compensation, types  B, E, J, K, N, R, S and T, and 1 measuring input for analog output signals 4...20 mA or 0...10 V.
    Order-Code: LTC-MP-3I (must be ordered together with the base unit)
    Extended scope of delivery:
    Electrical connection cables, plugs/terminals (red/black) for electrical connections

    Accuracy (max. deviation) of resistance and thermocouple measurement inputs with option LTC-MP-3I:
    Pt 100 or Pt 1000 at -40°C: ±0.09°C at 0°C: ±0.08°C at +150°C: ±0.11°C at +300°C: ±0.14°C
    Thermocouple type B at +950°C: ±0.97°C at +1050°C: ±1.03°C at +1200°C: ±1.12°C
    Thermocouple type E at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C
    Thermocouple type J at +200°C: ±0.52°C at +450°C: ±0.67°C at +700°C: ±0.82°C
    Thermocouple types K, N, R and S at +400°C: ±0.64°C at +700°C: ±0.82°C at +1000°C: ±1.00°C
    Thermocouple type T at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C

    Option LTC-MP-115V - device version for für 115 VAC

    The LTC-MP-115V option can be ordered together with the LR-Cal LTC-FB-9018-60:

    Device version for power supply 115 VAC 50/60 Hz (instead of für 230 VAC).

  • Specification

    LR-Cal LTC-FB-9018-60 specification

    Description: Specification:
    Temperature range: -18...+140°C
    Display + operation: color touchscreen
    Display accuracy: ±0.15°C *
    Measuring units: °C / °F / K
    Displey resolution: 0.1° / 0.01° / 0.00^°
    Mean heating time:
    (incl. stabilization)
    approx. 50 min.
    from ambient temperature to +140°C
    Mean cooling time:
    (incl. stabilization)
    approx. 50 min.
    from +140°C to ambient temperature
    Stability: ±0.02°C at 0°C **
    Axial uniformity: ±0.05°C
    Radial uniformity: ±0.07°C at +140°C ***
    Hole diameter: 60 mm
    Hole depth: 170 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4.5 VDC
    Adjustable ramp function: from 0.1°C/min
    Computer interface: USB type B
    Automatic calibration: at 5 temperature points
    External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
    Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric powert: 300 VA
    Dimensions: 160 x 360 x 350 mm
    Weight incl. standard accessories: 11.0 kg

    All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
    The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 40 mm.

  • Scope of standard delivery

    LR-Cal LTC-FB-9018-60 scope of delivery

    • Portable temperature calibration micro bath LR-Cal LTC-FB-9018-60
    • Spare fuses
    • Power supply cable
    • Connection cables for thermostat tests
    • Bottle with 500 cm³ silicone oil 47V10
    • Holding support for test items, e.g. glass thermometers
    • Sealing cap for the reservoir (for transport purposes)
    • Carrying bag with shoulder trap
    • Only if option LTC-MP-3I is ordered with the device: electrical connection cables and plugs (red/black) for the three measuring inputs
    • Acceptance test certficate 3.1 in accordance with DIN EN 10204 (factory calibration certificate)
    • Operating manual DE+EN
  • Options + Accessories

LR-Cal LTC-FB-9018-60 options + accessories

  • LTC-MP-3I

    3 configurable
    measuring inputs

    Code: LTC-MP-3I

    Option, must be ordered together with the calibrator:
    Device version with 3 configurable measuring inputs
    2 x for Pt 100 / Pt 1000 / Thermocouples
    1 x for analog signal 4-20mA / 0-10V

  • LTC-MP-115V

    Power supply
    115 VAC

    Code: LTC-MP-115V

    Option, must be ordered together with the calibrator:
    Device version for power supply
    115 VAC 50/60 Hz
    (instead of 230 VAC)
    NO extra charge

  • LTC-FB-INSERT

    Exchangeable
    calibration bath insert

    Code: LTC-FB-INSERT

    Option:
    Calibration bath insert incl. closing cap
    For fast change of the silicone oil resp. the calibration fluid (without empying and cleaning the reservoir)

  • LTC-FB-LONG

    Extension of the
    immersion depth

    Code: LTC-FB-LONG

    Accessory:
    Extension tube to increase the immersion depth
    Total length 255 mm
    Usable immersion depth 230 mm

  • LTC-FB-COOL


    Cooling coil

    Code: LTC-FB-COOL

    Accessory:
    Stainless steel cooling coil. The cooling coil shortens the cooling time, and it can also shift the start of the working range of the calibrator downwards (but above 0°C).
    Requires fresh water connection

  • LTC-AQ2SP2

    Windows-PC
    software Aq2Sp2

    Code: LTC-AQ2SP2

    Accessory:
    Windows-PC software Aq2Sp2
    (delivered as download link)

    Download Manual (PDF)

  • LTC-FB-DB-60-0

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-0

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    Without holes (for self-drilling)

  • LTC-FB-DB-60-9

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-9

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    With 9 drillings/holes: 4.0+4.0+4.5
    +5.5+6.5+6.5+8.5+10.5+12.5 mm

  • LTC-FB-DB-KOFFER

    Sturdy case
    Aluminum

    Code: LTC-FB-DB-KOFFER

    Accessory:
    Sturdy case made of aluminum with molded foam inserts
    Dimension 450 x 280 x 380 mm
    Weight 4.8 kg
    Shipping dimensions:
    600 x 370 x 500 mm

  • LTC-FB-DB-TASCHE

    Carrying bag
    with shoulder trap

    Code: LTC-FB-DB-TASCHE

    Spare part:
    Carrying bag with shoulder trap
    Dimension 410 x 220 x 380 mm
    Weight 1.45 kg
    Shipping dimensions:
    600 x 370 x 500 mm
    (1 bag ist included in calibrators delivery)

  • LTC-FB-ENT-TOOL

    Removal tools for
    protect. grill + mixer rod

    Code: LTC-FB-ENT-TOOL

    Accessory:
    Removal tools for protective grilles and magnetic mixer rod
    Facilitates emptying and cleaning
    (Already included in the scope of delivery of LTC-FB-INSERT and LTC-FB-DB-... inserts)

  • DKD-T-KAL-TK-6T

    DAkkS-
    certification

    Code: DKD-T-KAL-TK-6T

    Option:
    DAkkS certificate of calibration

    For the measuring chain consisting of display of the calibrator and reference sensor, 6 measuring points

  • LTC-FB-SO-010

    Silikone oil
    47V10

    Code: LTC-FB-SO-010

    Spare part:
    500 cm³ calibration fluid
    Silikone oil 47V10
    for temperature -30...+150°C
    (1 bottle included in calibrators scope of delivery)

  • LTC-FB-SO-020

    Silikone oil
    47V20

    Code: LTC-FB-SO-020

    Accessory:
    500 cm³ calibration fluid
    Silikone oil 47V20
    for temperature -20...+200°C


Suitable reference sensors (probes) for calibrator version with option LTC-MP-3I (3 configurable measuring inputs):

  • What is temperature calibration with a temperature calibration bath?

    What is temperature calibration with a temperature calibration bath?

    The test item (temperature measuring instruments of all types, e.g. dial thermometers, glass thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.

    Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.

    The LR-Cal temperature calibration baths have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.

    The LTC-MP-3I option is available for the portable LR-Cal LTC temperature calibration baths (with the exception of the LR-Cal LTC 150/250-F(-PLUS) devices), with three measuring inputs (2 x for RTD/TC and 1 x for analog signal 4...20 mA/0...10 V).

    The stationary LR-Cal temperature calibration baths are available in two versions: standard version and “-2I” version with 2 measuring inputs (for RTD/TC).

    Furthermore, the LR-Cal Aq2Sp2 PC software is available as an option for the LR-Cal temperature calibration baths. With this, the calibrator can be completely controlled from the PC via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.

    Synonyms

    Temperature calibration baths are also referred to as

    • Calibration bath
    • Micro bath
    • Temperature bath
    • Temperature calibrator
    • Temperature calibration device
    • Temperature calibration bath
  • Information about operating fluids for temperature calibration baths

    Calibration liquids for LR-Cal temperature calibration baths

    Applications:

    • Thermostatic fluids for micro baths and thermostatic temperature calibration baths

    Main Characteristics:

    • Operating range -50...+200°C
    • High stability
    • Good thermal conductivity
    • Nearly odourless
    • Long service live
    • Packaged in bottle or canister
    • Different types for a wide range applications
    • Low toxicity
    • Low viscosity and corrosovity

    Our thermostatic fluids are suitable for use in liquid thermostatic calibrators and thermostatic baths for applications from -50°C to +200°C.
    The choice of thermostatic fluid is the most important aspect for obtaining the best results with regard to operation and temperature control of calibrators and baths.
    These fluids have perfect thermodynamic characteristics to guarantee the best performance in the various fields of application, ensuring safe and reliable operation as well as optimum heat transfer efficiency.
    Silicone-based fluids are chemically inert substances that do not affect metals, are resistant to weathering and, when used correctly, the very low formation of cracking and oxidation ensures a long service life.

    Most of our thermostatic fluids are available in different containers: 500 cm³ bottles or 9 kg canisters (25 kg on request).

    Temperature application ranges of the various operating fluids:

    Type of calibration
    medium (liquid)
    Usable
    range
    (1)
    Temperature above which fume hood
    should be provided
    (2)
    Fume
    burning
    (3)
    A viscosity of 10 cst exists at a temperature of
    (4)
    Freezing temperature
    (5)
    Glycol 56% + water * -40...+90°C +70°C >110°C  +20°C -43°C
    Silicone oil 47V5 -40...+130°C +110°C +136°C  +5°C  -65°C
    Silicone oil 47V10 -30...+150°C +120°C +160°C  +20°C  -65°C
    Silicone oil 47V20 -20...+200°C ** +130°C +230°C  +60°C  -60°C
    Silicone oil 47V50 +30...+250°C ** +140°C +280°C  +160°C  -55°C
    Silicone oil 47V100 +50...+200°C ** +170°C >300°C  +200°C  -55°C
    Silicone oil 47V150 +50...+200°C ** +190°C >300°C  -/-  -55°C
    Note: Note text:
    * Mixture of 56% Glycol + 44% water
    ** In order to avoid gelling, it is recommended not to exceed a temperature of 200°C for a long time; even for fluids with wider operating ranges
    (1) Usable range
    The “usable range” of a calibration liquid is the temperature range in which it can be used under optimum conditions. The range may be limited by viscosity, flash points, freezing points, boiling points, evaporation rates, tendency to gel (or polymerize), etc.
    There is no liquid that covers extremely wide temperature ranges. Therefore, you should ideally use a separate calibration liquid for each temperature range. The recommended range is the one in which the viscosity is not optimal, but still reaches values that allow proper use.
    (2) Use of a fume hood
    The use of an exhaust hood when using temperature calibration baths prevents users from inhaling vapors from calibration liquids. It is best to use extraction devices near the access opening to the calibration bath. Vapors can otherwise settle on the mucous membranes of the eyes or on the skin, causing discomfort. Silicone oils can form benzene and formaldehyde when they decompose at high temperatures. The table shows the temperature above which an extractor hood should be used.
    (3) Flash point
    The table shows the temperature value at which the presence of an open flame can ignite vapours combustion.
    (4) Viscosity
    Viscosity is a measure of the flow resistance of a liquid. Viscosity increases with decreasing temperature. Kinetic viscosity is the ratio of absolute viscosity to density. This is measured in stokes and centistokes (cSt). 1 cSt = 1 stoke divided by 100. The higher the cSt, the more viscous or dense the liquid. Excessively viscous liquids place a strain on the stirring and pumping mechanisms of the calibration bath and do not transfer heat sufficiently. The table shows the temperature value below which the performance of the temperature calibration bath is no longer optimal. The equivalent value of water at 20°C is 1.1 cst.
    (5) Freezing point
    Silicone oils also change their state below a certain temperature by solidifying. The table shows the value at which the calibration fluid changes its state and solidifies.

    Other physical and chemical properties

    Specific heat
    The specific heat is the amount of heat required to raise or lower the temperature by 1 °C per kg. The specific heat, called CP, is expressed in Kcal/Kg °C. The energy required to heat or cool the calibration liquids depends on the amount of liquid contained in the calibrator in kg and its specific heat. A high specific heat value requires more energy to change the temperature, but offers greater temperature stability. The average CP of silicone fluids is 35 Kcal/Kg °C. The CP of water at 20 °C is equal to 1.

    Thermal conductivity
    Thermal conductivity is the ability of the liquid to transfer heat from one molecule to another. The better the heat transfer, the faster the liquid heats or cools. Better thermal conductivity improves the temperature uniformity in the reservoir of the calibration bath.

    Expansion
    All liquids have a coefficient of thermal expansion. This indicates how much the volume of a liquid changes (expands or contracts) when the temperature changes. Liquid expansion has an important impact on the safety, cleanliness and maintenance of temperature calibration baths. If calibrators are filled with too much liquid at a low temperature and then heated without regard to the increase in volume, they can overflow.

    If too little liquid is filled in the reservoir, below the level of the heating elements, the temperature rise of the non-immersed part of the heating elements could ignite the liquid.

    Gelation (polymerization)
    The gelling temperature is the temperature at which silicone fluids oxidize, gel and turn into a viscous slurry. The main cause of this phenomenon is oxidation. Although silicone fluids can be used safely up to their flash point, the susceptibility to polymerization increases when used above their oxidation point. To prevent polymerization, it is advisable to keep the operating time above the oxidation point of the silicone oil used as short as possible in order to keep contaminants (salts, other oils and oxidizing agents) away. The calibration fluid must be replaced if it becomes dark, viscous or temperature unstable. It is important to limit the application time at high temperatures to prevent the oil from degrading and to extend its service life.

Read more …LR-Cal LTC-FB-9018-60 Temperature micro bath -18...+140°C

  • Hits: 507

LR-Cal LTC-FB-0200-60 Temperature micro bath ...+200°C

LR-Cal LTC-FB-0200-60
Temperature calibration micro bath ...+200°C

Brochure

LR-Cal LTC-FB-0200-60 temperature calibration micro bath LR-Cal
  • Portable
  • Temperature range ambient temperature...+200°C
  • Accuracy ±0.2°C, stability ±0.03°C
  • Bath volume approx. 500 cm³
    (60 mm diameter, 170 mm depth)
  • Option: changeable calibration bath insert
  • Temperature calibration micro bath LR-Cal LTC-FB-0200-60 with support for test item

    FLUID

    incl. support for test item and reference

  • Temperature calibration microbath LR-Cal LTC-FB-0200-60 touchscreen

    FLUID

    Simple operation via touchscreen

  • Exchangeable calibration bath insert for LR-Cal LTC-FB-0200-60

    FLUID

    LTC-FB-INSERT
    Changeable calibration bath insert

  • LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL
    Optional tool set for removing protective grilles + magnetic mixer rod (easier emptying+cleaning)

  • Description

    LR-Cal LTC-FB-0200-60 Description

    The LR-Cal LTC-FB-0200-60 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

    • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
    • Homogeneous temperature distribution due to adjustable magnetic mixer
    • Simple operation with color touchscreen
    • Ramp function + thermostat tests
    • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
    • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
    • Option: dry block inserts to convert the device to a dry block temperature calibrator
  • Options LTC-MP-3I (3 measuring inputs) + LTC-MP-115V (for 115 VAC)

    Option LTC-MP-3I - three configurable measuring inputs

    Option LTC-MP-3I: 3 konfigurierbare MesseingängeThe LTC-MP-3I option can be ordered together with the LR-Cal LTC-FB-0200-60:

    Device version with 3 configurable measuring inputs. Of these, 2 for resistance thermometers (Pt 100 (2-, 3- or 4-wire) or Pt 100, or thermocouples incl. cold junction compensation, types  B, E, J, K, N, R, S and T, and 1 measuring input for analog output signals 4...20 mA or 0...10 V.
    Order-Code: LTC-MP-3I (must be ordered together with the base unit)
    Extended scope of delivery:
    Electrical connection cables, plugs/terminals (red/black) for electrical connections

    Accuracy (max. deviation) of resistance and thermocouple measurement inputs with option LTC-MP-3I:
    Pt 100 or Pt 1000 at -40°C: ±0.09°C at 0°C: ±0.08°C at +150°C: ±0.11°C at +300°C: ±0.14°C
    Thermocouple type B at +950°C: ±0.97°C at +1050°C: ±1.03°C at +1200°C: ±1.12°C
    Thermocouple type E at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C
    Thermocouple type J at +200°C: ±0.52°C at +450°C: ±0.67°C at +700°C: ±0.82°C
    Thermocouple types K, N, R and S at +400°C: ±0.64°C at +700°C: ±0.82°C at +1000°C: ±1.00°C
    Thermocouple type T at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C

    Option LTC-MP-115V - device version for für 115 VAC

    The LTC-MP-115V option can be ordered together with the LR-Cal LTC-FB-0200-60:

    Device version for power supply 115 VAC 50/60 Hz (instead of für 230 VAC).

  • Specification

    LR-Cal LTC-FB-0200-60 specification

    Description: Specification:
    Temperature range: ambient temperature...+200°C
    Display + operation: color touchscreen
    Display accuracy: ±0.2°C *
    Measuring units: °C / °F / K
    Displey resolution: 0.1° / 0.01° / 0.00^°
    Mean heating time:
    (incl. stabilization)
    approx. 30 min.
    from ambient temperature to +200°C
    Mean cooling time:
    (incl. stabilization)
    approx. 75 min.
    from +200°C to ambient temperature
    Stability: ±0.03°C at full temperature range **
    Axial uniformity: ±0.01°C at +140°C
    Radial uniformity: ±0.06°C at +140°C ***
    Hole diameter: 60 mm
    Hole depth: 170 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4.5 VDC
    Adjustable ramp function: from 0.1°C/min
    Computer interface: USB type B
    Automatic calibration: at 5 temperature points
    External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
    Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric powert: 500 VA
    Dimensions: 160 x 360 x 350 mm
    Weight incl. standard accessories: 8.3 kg

    All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
    The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 40 mm.

  • Scope of standard delivery

    LR-Cal LTC-FB-0200-60 scope of delivery

    • Portable temperature calibration micro bath LR-Cal LTC-FB-0200-60
    • Spare fuses
    • Power supply cable
    • Connection cables for thermostat tests
    • Bottle with 500 cm³ silicone oil 47V20
    • Holding support for test items, e.g. glass thermometers
    • Sealing cap for the reservoir (for transport purposes)
    • Carrying bag with shoulder trap
    • Only if option LTC-MP-3I is ordered with the device: electrical connection cables and plugs (red/black) for the three measuring inputs
    • Acceptance test certficate 3.1 in accordance with DIN EN 10204 (factory calibration certificate)
    • Operating manual DE+EN
  • Options + Accessories

LR-Cal LTC-FB-0200-60 options + accessories

  • LTC-MP-3I

    3 configurable
    measuring inputs

    Code: LTC-MP-3I

    Option, must be ordered together with the calibrator:
    Device version with 3 configurable measuring inputs
    2 x for Pt 100 / Pt 1000 / Thermocouples
    1 x for analog signal 4-20mA / 0-10V

  • LTC-MP-115V

    Power supply
    115 VAC

    Code: LTC-MP-115V

    Option, must be ordered together with the calibrator:
    Device version for power supply
    115 VAC 50/60 Hz
    (instead of 230 VAC)
    NO extra charge

  • LTC-FB-INSERT

    Exchangeable
    calibration bath insert

    Code: LTC-FB-INSERT

    Option:
    Calibration bath insert incl. closing cap
    For fast change of the silicone oil resp. the calibration fluid (without empying and cleaning the reservoir)

  • LTC-FB-LONG

    Extension of the
    immersion depth

    Code: LTC-FB-LONG

    Accessory:
    Extension tube to increase the immersion depth
    Total length 255 mm
    Usable immersion depth 230 mm

  • LTC-FB-COOL


    Cooling coil

    Code: LTC-FB-COOL

    Accessory:
    Stainless steel cooling coil. The cooling coil shortens the cooling time, and it can also shift the start of the working range of the calibrator downwards (but above 0°C).
    Requires fresh water connection

  • LTC-AQ2SP2

    Windows-PC
    software Aq2Sp2

    Code: LTC-AQ2SP2

    Accessory:
    Windows-PC software Aq2Sp2
    (delivered as download link)

    Download Manual (PDF)

  • LTC-FB-DB-60-0

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-0

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    Without holes (for self-drilling)

  • LTC-FB-DB-60-9

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-9

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    With 9 drillings/holes: 4.0+4.0+4.5
    +5.5+6.5+6.5+8.5+10.5+12.5 mm

  • LTC-FB-DB-KOFFER

    Sturdy case
    Aluminum

    Code: LTC-FB-DB-KOFFER

    Accessory:
    Sturdy case made of aluminum with molded foam inserts
    Dimension 450 x 280 x 380 mm
    Weight 4.8 kg
    Shipping dimensions:
    600 x 370 x 500 mm

  • LTC-FB-DB-TASCHE

    Carrying bag
    with shoulder trap

    Code: LTC-FB-DB-TASCHE

    Spare part:
    Carrying bag with shoulder trap
    Dimension 410 x 220 x 380 mm
    Weight 1.45 kg
    Shipping dimensions:
    600 x 370 x 500 mm
    (1 bag ist included in calibrators delivery)

  • LTC-FB-ENT-TOOL

    Removal tools for
    protect. grill + mixer rod

    Code: LTC-FB-ENT-TOOL

    Accessory:
    Removal tools for protective grilles and magnetic mixer rod
    Facilitates emptying and cleaning
    (Already included in the scope of delivery of LTC-FB-INSERT and LTC-FB-DB-... inserts)

  • DKD-T-KAL-TK-6T

    DAkkS-
    certification

    Code: DKD-T-KAL-TK-6T

    Option:
    DAkkS certificate of calibration

    For the measuring chain consisting of display of the calibrator and reference sensor, 6 measuring points

  • LTC-FB-SO-020

    Silikone oil
    47V20

    Code: LTC-FB-SO-020

    Spare part:
    500 cm³ calibration fluid
    Silikone oil 47V20
    for temperature -20...+200°C
    (1 bottle included in calibrators scope of delivery)

  • LTC-FB-SO-050

    Silikone oil
    47V50

    Code: LTC-FB-SO-050

    Accessory:
    500 cm³ calibration fluid
    Silikone oil 47V50
    for temperature +30...+200°C

  • LTC-FB-SO-100

    Silikone oil
    47V100

    Code: LTC-FB-SO-100

    Accessory:
    500 cm³ calibration fluid
    Silikone oil 47V100
    for temperature +50...+200°C


Suitable reference sensors (probes) for calibrator version with option LTC-MP-3I (3 configurable measuring inputs):

  • What is temperature calibration with a temperature calibration bath?

    What is temperature calibration with a temperature calibration bath?

    The test item (temperature measuring instruments of all types, e.g. dial thermometers, glass thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.

    Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.

    The LR-Cal temperature calibration baths have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.

    The LTC-MP-3I option is available for the portable LR-Cal LTC temperature calibration baths (with the exception of the LR-Cal LTC 150/250-F(-PLUS) devices), with three measuring inputs (2 x for RTD/TC and 1 x for analog signal 4...20 mA/0...10 V).

    The stationary LR-Cal temperature calibration baths are available in two versions: standard version and “-2I” version with 2 measuring inputs (for RTD/TC).

    Furthermore, the LR-Cal Aq2Sp2 PC software is available as an option for the LR-Cal temperature calibration baths. With this, the calibrator can be completely controlled from the PC via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.

    Synonyms

    Temperature calibration baths are also referred to as

    • Calibration bath
    • Micro bath
    • Temperature bath
    • Temperature calibrator
    • Temperature calibration device
    • Temperature calibration bath
  • Information about operating fluids for temperature calibration baths

    Calibration liquids for LR-Cal temperature calibration baths

    Applications:

    • Thermostatic fluids for micro baths and thermostatic temperature calibration baths

    Main Characteristics:

    • Operating range -50...+200°C
    • High stability
    • Good thermal conductivity
    • Nearly odourless
    • Long service live
    • Packaged in bottle or canister
    • Different types for a wide range applications
    • Low toxicity
    • Low viscosity and corrosovity

    Our thermostatic fluids are suitable for use in liquid thermostatic calibrators and thermostatic baths for applications from -50°C to +200°C.
    The choice of thermostatic fluid is the most important aspect for obtaining the best results with regard to operation and temperature control of calibrators and baths.
    These fluids have perfect thermodynamic characteristics to guarantee the best performance in the various fields of application, ensuring safe and reliable operation as well as optimum heat transfer efficiency.
    Silicone-based fluids are chemically inert substances that do not affect metals, are resistant to weathering and, when used correctly, the very low formation of cracking and oxidation ensures a long service life.

    Most of our thermostatic fluids are available in different containers: 500 cm³ bottles or 9 kg canisters (25 kg on request).

    Temperature application ranges of the various operating fluids:

    Type of calibration
    medium (liquid)
    Usable
    range
    (1)
    Temperature above which fume hood
    should be provided
    (2)
    Fume
    burning
    (3)
    A viscosity of 10 cst exists at a temperature of
    (4)
    Freezing temperature
    (5)
    Glycol 56% + water * -40...+90°C +70°C >110°C  +20°C -43°C
    Silicone oil 47V5 -40...+130°C +110°C +136°C  +5°C  -65°C
    Silicone oil 47V10 -30...+150°C +120°C +160°C  +20°C  -65°C
    Silicone oil 47V20 -20...+200°C ** +130°C +230°C  +60°C  -60°C
    Silicone oil 47V50 +30...+250°C ** +140°C +280°C  +160°C  -55°C
    Silicone oil 47V100 +50...+200°C ** +170°C >300°C  +200°C  -55°C
    Silicone oil 47V150 +50...+200°C ** +190°C >300°C  -/-  -55°C
    Note: Note text:
    * Mixture of 56% Glycol + 44% water
    ** In order to avoid gelling, it is recommended not to exceed a temperature of 200°C for a long time; even for fluids with wider operating ranges
    (1) Usable range
    The “usable range” of a calibration liquid is the temperature range in which it can be used under optimum conditions. The range may be limited by viscosity, flash points, freezing points, boiling points, evaporation rates, tendency to gel (or polymerize), etc.
    There is no liquid that covers extremely wide temperature ranges. Therefore, you should ideally use a separate calibration liquid for each temperature range. The recommended range is the one in which the viscosity is not optimal, but still reaches values that allow proper use.
    (2) Use of a fume hood
    The use of an exhaust hood when using temperature calibration baths prevents users from inhaling vapors from calibration liquids. It is best to use extraction devices near the access opening to the calibration bath. Vapors can otherwise settle on the mucous membranes of the eyes or on the skin, causing discomfort. Silicone oils can form benzene and formaldehyde when they decompose at high temperatures. The table shows the temperature above which an extractor hood should be used.
    (3) Flash point
    The table shows the temperature value at which the presence of an open flame can ignite vapours combustion.
    (4) Viscosity
    Viscosity is a measure of the flow resistance of a liquid. Viscosity increases with decreasing temperature. Kinetic viscosity is the ratio of absolute viscosity to density. This is measured in stokes and centistokes (cSt). 1 cSt = 1 stoke divided by 100. The higher the cSt, the more viscous or dense the liquid. Excessively viscous liquids place a strain on the stirring and pumping mechanisms of the calibration bath and do not transfer heat sufficiently. The table shows the temperature value below which the performance of the temperature calibration bath is no longer optimal. The equivalent value of water at 20°C is 1.1 cst.
    (5) Freezing point
    Silicone oils also change their state below a certain temperature by solidifying. The table shows the value at which the calibration fluid changes its state and solidifies.

    Other physical and chemical properties

    Specific heat
    The specific heat is the amount of heat required to raise or lower the temperature by 1 °C per kg. The specific heat, called CP, is expressed in Kcal/Kg °C. The energy required to heat or cool the calibration liquids depends on the amount of liquid contained in the calibrator in kg and its specific heat. A high specific heat value requires more energy to change the temperature, but offers greater temperature stability. The average CP of silicone fluids is 35 Kcal/Kg °C. The CP of water at 20 °C is equal to 1.

    Thermal conductivity
    Thermal conductivity is the ability of the liquid to transfer heat from one molecule to another. The better the heat transfer, the faster the liquid heats or cools. Better thermal conductivity improves the temperature uniformity in the reservoir of the calibration bath.

    Expansion
    All liquids have a coefficient of thermal expansion. This indicates how much the volume of a liquid changes (expands or contracts) when the temperature changes. Liquid expansion has an important impact on the safety, cleanliness and maintenance of temperature calibration baths. If calibrators are filled with too much liquid at a low temperature and then heated without regard to the increase in volume, they can overflow.

    If too little liquid is filled in the reservoir, below the level of the heating elements, the temperature rise of the non-immersed part of the heating elements could ignite the liquid.

    Gelation (polymerization)
    The gelling temperature is the temperature at which silicone fluids oxidize, gel and turn into a viscous slurry. The main cause of this phenomenon is oxidation. Although silicone fluids can be used safely up to their flash point, the susceptibility to polymerization increases when used above their oxidation point. To prevent polymerization, it is advisable to keep the operating time above the oxidation point of the silicone oil used as short as possible in order to keep contaminants (salts, other oils and oxidizing agents) away. The calibration fluid must be replaced if it becomes dark, viscous or temperature unstable. It is important to limit the application time at high temperatures to prevent the oil from degrading and to extend its service life.

Read more …LR-Cal LTC-FB-0200-60 Temperature micro bath...+200°C

  • Hits: 366

LR-Cal LTC-FB-0250-60 Temperature micro bath ...+250°C

LR-Cal LTC-FB-0250-60
Temperature calibration micro bath ...+250°C

Brochure

LR-Cal LTC-FB-0250-60 temperature calibration micro bath LR-Cal
  • Portable
  • Temperature range ambient temperature...+250°C
  • Accuracy ±0.2°C, stability ±0.03°C
  • Bath volume approx. 500 cm³
    (60 mm diameter, 170 mm depth)
  • Option: changeable calibration bath insert
  • Temperature calibration micro bath LR-Cal LTC-FB-0250-60 with support for test item

    FLUID

    incl. support for test item and reference

  • Temperature calibration microbath LR-Cal LTC-FB-0250-60 touchscreen

    FLUID

    Simple operation via touchscreen

  • Exchangeable calibration bath insert for LR-Cal LTC-FB-0250-60

    FLUID

    LTC-FB-INSERT
    Changeable calibration bath insert

  • LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL

    LTC-FB-ENT-TOOL
    Optional tool set for removing protective grilles + magnetic mixer rod (easier emptying+cleaning)

  • Description

    LR-Cal LTC-FB-0250-60 Description

    The LR-Cal LTC-FB-0250-60 portable temperature calibration bath serves as a temperature source and reference instrument in one, for testing and calibrating all types of temperature measuring instruments. Various metal block inserts are also available.

    • Optimal for calibration of glass thermometers and laboratory thermometers as well as temperature sensors with not exactly cylindrical shape
    • Homogeneous temperature distribution due to adjustable magnetic mixer
    • Simple operation with color touchscreen
    • Ramp function + thermostat tests
    • Computer-controlled calibrations via USB interface and optional software LR-Cal Aq2Sp2 for MS-Windows® computers
    • Option: changeable calibration bath insert for changing the calibration fluid without emptying and cleaning of the reservoir
    • Option: dry block inserts to convert the device to a dry block temperature calibrator
  • Options LTC-MP-3I (3 measuring inputs) + LTC-MP-115V (for 115 VAC)

    Option LTC-MP-3I - three configurable measuring inputs

    Option LTC-MP-3I: 3 konfigurierbare MesseingängeThe LTC-MP-3I option can be ordered together with the LR-Cal LTC-FB-0250-60:

    Device version with 3 configurable measuring inputs. Of these, 2 for resistance thermometers (Pt 100 (2-, 3- or 4-wire) or Pt 100, or thermocouples incl. cold junction compensation, types  B, E, J, K, N, R, S and T, and 1 measuring input for analog output signals 4...20 mA or 0...10 V.
    Order-Code: LTC-MP-3I (must be ordered together with the base unit)
    Extended scope of delivery:
    Electrical connection cables, plugs/terminals (red/black) for electrical connections

    Accuracy (max. deviation) of resistance and thermocouple measurement inputs with option LTC-MP-3I:
    Pt 100 or Pt 1000 at -40°C: ±0.09°C at 0°C: ±0.08°C at +150°C: ±0.11°C at +300°C: ±0.14°C
    Thermocouple type B at +950°C: ±0.97°C at +1050°C: ±1.03°C at +1200°C: ±1.12°C
    Thermocouple type E at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C
    Thermocouple type J at +200°C: ±0.52°C at +450°C: ±0.67°C at +700°C: ±0.82°C
    Thermocouple types K, N, R and S at +400°C: ±0.64°C at +700°C: ±0.82°C at +1000°C: ±1.00°C
    Thermocouple type T at -40°C: ±0.42°C at 0°C: ±0.40°C at +350°C: ±0.61°C

    Option LTC-MP-115V - device version for für 115 VAC

    The LTC-MP-115V option can be ordered together with the LR-Cal LTC-FB-0250-60:

    Device version for power supply 115 VAC 50/60 Hz (instead of für 230 VAC).

  • Specification

    LR-Cal LTC-FB-0250-60 specification

    Description: Specification:
    Temperature range: ambient temperature...+250°C
    Display + operation: color touchscreen
    Display accuracy: ±0.2°C *
    Measuring units: °C / °F / K
    Displey resolution: 0.1° / 0.01° / 0.00^°
    Mean heating time:
    (incl. stabilization)
    approx. 30 min.
    from ambient temperature to +200°C
    Mean cooling time:
    (incl. stabilization)
    approx. 75 min.
    from +200°C to ambient temperature
    Stability: ±0.03°C at full temperature range **
    Axial uniformity: ±0.01°C at +140°C
    Radial uniformity: ±0.06°C at +140°C ***
    Hole diameter: 60 mm
    Hole depth: 170 mm
    Insert material: anticorodal
    Switch test; voltage: Open/Closed; 4.5 VDC
    Adjustable ramp function: from 0.1°C/min
    Computer interface: USB type B
    Automatic calibration: at 5 temperature points
    External sensors/probes: Optional with Order-Code LTC-MP-3I (to be ordered together with instrument)
    Operation voltage: 230 VAC (optional 115 VAC) 50/60 Hz
    Electric powert: 500 VA
    Dimensions: 160 x 360 x 350 mm
    Weight incl. standard accessories: 8.3 kg

    All values have been measured at ambient temperature of +20°C and with a sensor with diamter 6 mm.
    The achievement of stabilization is confirmec by a displayed root symbol and an audial beep.
    * = Temperature deviation between display and reference probe.
    ** = Max. temperature difference at stable temperature over 30 min.
    *** = Measured from the button up to 40 mm.

  • Scope of standard delivery

    LR-Cal LTC-FB-0250-60 scope of delivery

    • Portable temperature calibration micro bath LR-Cal LTC-FB-0250-60
    • Spare fuses
    • Power supply cable
    • Connection cables for thermostat tests
    • Bottle with 500 cm³ silicone oil 47V100
    • Holding support for test items, e.g. glass thermometers
    • Sealing cap for the reservoir (for transport purposes)
    • Carrying bag with shoulder trap
    • Only if option LTC-MP-3I is ordered with the device: electrical connection cables and plugs (red/black) for the three measuring inputs
    • Acceptance test certficate 3.1 in accordance with DIN EN 10204 (factory calibration certificate)
    • Operating manual DE+EN
  • Options + Accessories

LR-Cal LTC-FB-0250-60 options + accessories

  • LTC-MP-3I

    3 configurable
    measuring inputs

    Code: LTC-MP-3I

    Option, must be ordered together with the calibrator:
    Device version with 3 configurable measuring inputs
    2 x for Pt 100 / Pt 1000 / Thermocouples
    1 x for analog signal 4-20mA / 0-10V

  • LTC-MP-115V

    Power supply
    115 VAC

    Code: LTC-MP-115V

    Option, must be ordered together with the calibrator:
    Device version for power supply
    115 VAC 50/60 Hz
    (instead of 230 VAC)
    NO extra charge

  • LTC-FB-INSERT

    Exchangeable
    calibration bath insert

    Code: LTC-FB-INSERT

    Option:
    Calibration bath insert incl. closing cap
    For fast change of the silicone oil resp. the calibration fluid (without empying and cleaning the reservoir)

  • LTC-FB-LONG

    Extension of the
    immersion depth

    Code: LTC-FB-LONG

    Accessory:
    Extension tube to increase the immersion depth
    Total length 255 mm
    Usable immersion depth 230 mm

  • LTC-FB-COOL


    Cooling coil

    Code: LTC-FB-COOL

    Accessory:
    Stainless steel cooling coil. The cooling coil shortens the cooling time, and it can also shift the start of the working range of the calibrator downwards (but above 0°C).
    Requires fresh water connection

  • LTC-AQ2SP2

    Windows-PC
    software Aq2Sp2

    Code: LTC-AQ2SP2

    Accessory:
    Windows-PC software Aq2Sp2
    (delivered as download link)

    Download Manual (PDF)

  • LTC-FB-DB-60-0

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-0

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    Without holes (for self-drilling)

  • LTC-FB-DB-60-9

    Conversion kit to
    dry block calibrator

    Code: LTC-FB-DB-60-9

    Accessory:
    Dry block conversion kit (aluminum)
    Diameter 60 mm
    Depth 170 mm
    With 9 drillings/holes: 4.0+4.0+4.5
    +5.5+6.5+6.5+8.5+10.5+12.5 mm

  •  LTC-FB-DB-KOFFER

    Sturdy case
    Aluminum

    Code: LTC-FB-DB-KOFFER

    Accessory:
    Sturdy case made of aluminum with molded foam inserts
    Dimension 450 x 280 x 380 mm
    Weight 4.8 kg
    Shipping dimensions:
    600 x 370 x 500 mm

  • LTC-FB-DB-TASCHE

    Carrying bag
    with shoulder trap

    Code: LTC-FB-DB-TASCHE

    Spare part:
    Carrying bag with shoulder trap
    Dimension 410 x 220 x 380 mm
    Weight 1.45 kg
    Shipping dimensions:
    600 x 370 x 500 mm
    (1 bag ist included in calibrators delivery)

  • LTC-FB-ENT-TOOL

    Removal tools for
    protect. grill + mixer rod

    Code: LTC-FB-ENT-TOOL

    Accessory:
    Removal tools for protective grilles and magnetic mixer rod
    Facilitates emptying and cleaning
    (Already included in the scope of delivery of LTC-FB-INSERT and LTC-FB-DB-... inserts)

  • DKD-T-KAL-TK-6T

    DAkkS-
    certification

    Code: DKD-T-KAL-TK-6T

    Option:
    DAkkS certificate of calibration

    For the measuring chain consisting of display of the calibrator and reference sensor, 6 measuring points

  • LTC-FB-SO-050

    Silikone oil
    47V50

    Code: LTC-FB-SO-050

    Accessory:
    500 cm³ calibration fluid
    Silikone oil 47V50
    for temperature +30...+200°C

  • LTC-FB-SO-100

    Silikone oil
    47V100

    Code: LTC-FB-SO-100

    Spare part:
    500 cm³ calibration fluid
    Silikone oil 47V100
    for temperature +50...+200°C
    (1 bottle included in calibrators scope of delivery)

  • LTC-FB-SO-150

    Silikone oil
    47V150

    Code: LTC-FB-SO-150

    Accessory:
    500 cm³ calibration fluid
    Silikone oil 47V150
    for temperature +50...+200°C


Suitable reference sensors (probes) for calibrator version with option LTC-MP-3I (3 configurable measuring inputs):

  • What is temperature calibration with a temperature calibration bath?

    What is temperature calibration with a temperature calibration bath?

    The test item (temperature measuring instruments of all types, e.g. dial thermometers, glass thermometers, resistance thermometers, temperature switches, thermostats, etc.) and the reference temperature measuring instrument (installed in the calibrator or external reference) are both subjected to test temperatures. The measured values of the reference are compared with the values of the test item. In this way, a statement can be made not only about the function of the test item, but also about the measurement accuracy or the measured value deviation of the test item.

    Please note that the specifications regarding radial and axial temperature uniformity given in the respective technical data refer to specific immersion depths and do not necessarily apply to all immersion depths in every case. The temperature range specifications apply at an ambient temperature of +20°C.

    The LR-Cal temperature calibration baths have an internal (built-in) reference. If its accuracy is not sufficient, an external reference (without or with its own display) can be used.

    The LTC-MP-3I option is available for the portable LR-Cal LTC temperature calibration baths (with the exception of the LR-Cal LTC 150/250-F(-PLUS) devices), with three measuring inputs (2 x for RTD/TC and 1 x for analog signal 4...20 mA/0...10 V).

    The stationary LR-Cal temperature calibration baths are available in two versions: standard version and “-2I” version with 2 measuring inputs (for RTD/TC).

    Furthermore, the LR-Cal Aq2Sp2 PC software is available as an option for the LR-Cal temperature calibration baths. With this, the calibrator can be completely controlled from the PC via its serial interface, manual or automatic calibration of one or more DUTs, load and lifetime tests, thermostat tests, as well as the creation of calibration certificates.

    Synonyms

    Temperature calibration baths are also referred to as

    • Calibration bath
    • Micro bath
    • Temperature bath
    • Temperature calibrator
    • Temperature calibration device
    • Temperature calibration bath
  • Information about operating fluids for temperature calibration baths

    Calibration liquids for LR-Cal temperature calibration baths

    Applications:

    • Thermostatic fluids for micro baths and thermostatic temperature calibration baths

    Main Characteristics:

    • Operating range -50...+200°C
    • High stability
    • Good thermal conductivity
    • Nearly odourless
    • Long service live
    • Packaged in bottle or canister
    • Different types for a wide range applications
    • Low toxicity
    • Low viscosity and corrosovity

    Our thermostatic fluids are suitable for use in liquid thermostatic calibrators and thermostatic baths for applications from -50°C to +200°C.
    The choice of thermostatic fluid is the most important aspect for obtaining the best results with regard to operation and temperature control of calibrators and baths.
    These fluids have perfect thermodynamic characteristics to guarantee the best performance in the various fields of application, ensuring safe and reliable operation as well as optimum heat transfer efficiency.
    Silicone-based fluids are chemically inert substances that do not affect metals, are resistant to weathering and, when used correctly, the very low formation of cracking and oxidation ensures a long service life.

    Most of our thermostatic fluids are available in different containers: 500 cm³ bottles or 9 kg canisters (25 kg on request).

    Temperature application ranges of the various operating fluids:

    Type of calibration
    medium (liquid)
    Usable
    range
    (1)
    Temperature above which fume hood
    should be provided
    (2)
    Fume
    burning
    (3)
    A viscosity of 10 cst exists at a temperature of
    (4)
    Freezing temperature
    (5)
    Glycol 56% + water * -40...+90°C +70°C >110°C  +20°C -43°C
    Silicone oil 47V5 -40...+130°C +110°C +136°C  +5°C  -65°C
    Silicone oil 47V10 -30...+150°C +120°C +160°C  +20°C  -65°C
    Silicone oil 47V20 -20...+200°C ** +130°C +230°C  +60°C  -60°C
    Silicone oil 47V50 +30...+250°C ** +140°C +280°C  +160°C  -55°C
    Silicone oil 47V100 +50...+200°C ** +170°C >300°C  +200°C  -55°C
    Silicone oil 47V150 +50...+200°C ** +190°C >300°C  -/-  -55°C
    Note: Note text:
    * Mixture of 56% Glycol + 44% water
    ** In order to avoid gelling, it is recommended not to exceed a temperature of 200°C for a long time; even for fluids with wider operating ranges
    (1) Usable range
    The “usable range” of a calibration liquid is the temperature range in which it can be used under optimum conditions. The range may be limited by viscosity, flash points, freezing points, boiling points, evaporation rates, tendency to gel (or polymerize), etc.
    There is no liquid that covers extremely wide temperature ranges. Therefore, you should ideally use a separate calibration liquid for each temperature range. The recommended range is the one in which the viscosity is not optimal, but still reaches values that allow proper use.
    (2) Use of a fume hood
    The use of an exhaust hood when using temperature calibration baths prevents users from inhaling vapors from calibration liquids. It is best to use extraction devices near the access opening to the calibration bath. Vapors can otherwise settle on the mucous membranes of the eyes or on the skin, causing discomfort. Silicone oils can form benzene and formaldehyde when they decompose at high temperatures. The table shows the temperature above which an extractor hood should be used.
    (3) Flash point
    The table shows the temperature value at which the presence of an open flame can ignite vapours combustion.
    (4) Viscosity
    Viscosity is a measure of the flow resistance of a liquid. Viscosity increases with decreasing temperature. Kinetic viscosity is the ratio of absolute viscosity to density. This is measured in stokes and centistokes (cSt). 1 cSt = 1 stoke divided by 100. The higher the cSt, the more viscous or dense the liquid. Excessively viscous liquids place a strain on the stirring and pumping mechanisms of the calibration bath and do not transfer heat sufficiently. The table shows the temperature value below which the performance of the temperature calibration bath is no longer optimal. The equivalent value of water at 20°C is 1.1 cst.
    (5) Freezing point
    Silicone oils also change their state below a certain temperature by solidifying. The table shows the value at which the calibration fluid changes its state and solidifies.

    Other physical and chemical properties

    Specific heat
    The specific heat is the amount of heat required to raise or lower the temperature by 1 °C per kg. The specific heat, called CP, is expressed in Kcal/Kg °C. The energy required to heat or cool the calibration liquids depends on the amount of liquid contained in the calibrator in kg and its specific heat. A high specific heat value requires more energy to change the temperature, but offers greater temperature stability. The average CP of silicone fluids is 35 Kcal/Kg °C. The CP of water at 20 °C is equal to 1.

    Thermal conductivity
    Thermal conductivity is the ability of the liquid to transfer heat from one molecule to another. The better the heat transfer, the faster the liquid heats or cools. Better thermal conductivity improves the temperature uniformity in the reservoir of the calibration bath.

    Expansion
    All liquids have a coefficient of thermal expansion. This indicates how much the volume of a liquid changes (expands or contracts) when the temperature changes. Liquid expansion has an important impact on the safety, cleanliness and maintenance of temperature calibration baths. If calibrators are filled with too much liquid at a low temperature and then heated without regard to the increase in volume, they can overflow.

    If too little liquid is filled in the reservoir, below the level of the heating elements, the temperature rise of the non-immersed part of the heating elements could ignite the liquid.

    Gelation (polymerization)
    The gelling temperature is the temperature at which silicone fluids oxidize, gel and turn into a viscous slurry. The main cause of this phenomenon is oxidation. Although silicone fluids can be used safely up to their flash point, the susceptibility to polymerization increases when used above their oxidation point. To prevent polymerization, it is advisable to keep the operating time above the oxidation point of the silicone oil used as short as possible in order to keep contaminants (salts, other oils and oxidizing agents) away. The calibration fluid must be replaced if it becomes dark, viscous or temperature unstable. It is important to limit the application time at high temperatures to prevent the oil from degrading and to extend its service life.

Read more …LR-Cal LTC-FB-0250-60 Temperature micro bath...+250°C

  • Hits: 624