Calibration Hand Pump
LPP 1000
Portable Pressure Test System |
 |
|

LPP 1000: Calibration
Hand Pump,
1000 bar (14500 psi)
Portable Pressure Test System
High pressure test hose included!

Data Sheet: 
Operating Manual: 
|
The
LPP 1000 hand held test pump is
lightweight, easy to use and can generate pressures up to 1000 bar / 14500 psi.
The
LPP 1000 is well suited for calibrating pressure
transmitters, pressure transducers, pressure switches and pressure gauges.
The
LPP 1000 can be used as a Portable Pressure
Comparator in conjunction with various pressure
indicating devices. The system offers adjustable over pressure
protection, fine control, an a unique selector switch to move from
priming to high pressure generation.
Features:
- Hydraulic Pressure up to 1000 bar / 14500 psi!
- Sealed Construction - no leaks - no rust!
- Including flexible hose for connection to
test specimen
- Lightweight, hand held and easy to use.
- Quick priming and pressure generation using scissor
action pump.
- High pressure / priming selector.
- Fine adjustment control.
- Pressure release valve.
- Large see-through reservoir with a large capacitiy
- Suitable for mineral oil based hydraulic fluids or clean water,
free of calcium-carbonate / scale as operation fluid.
(Do not use water-based hydraulic oils or
destilled water)
- Skydrol- and brake fluid version also
available ("-S".
- Optional accessories include a carrying case, a set
of adaptors suitable for many applications.
Specification:
| Pressure
range: |
0 to
14500 psi / 1000 bar |
| Pressure
media: |
Mineral oil based hydraulic
fluids or
clean water, free of calcium-carbonate / scale
(Do NOT use water based hydraulic oils,
do NOT use destilled water)
Optional:
Skydrol or Brake Fluid (Mod. LPP 1000-S) |
| Pressure
connections: |
for master
instrument: 1/2" BSP female rotating
for unit under test: 1/4" BSP female at flexible hose |
|
Master Instruments available to suit
LPP 1000:
-
LPC 300 - Electronic Process
Calibrator, several ranges, accuracy 0.025% F.S.
-
LPC 200 - Electronic Pressure
Calibrator, several ranges, accuracy 0.025% F.S. / ±0.1% of reading
-
TLDMM - Electronic Pressure
Calibrator, accuracy 0.05% F.S.
-
TLDMM-A01 - Digital Test
Pressure Gauge, accuracy 0.1% f.s.
-
TLDMM-A02 - Digital Test
Pressure Gauge, accuracy 0.2% f.s.
-
DFP - Digital Reference Pressure
Gauge, accuracy 0.1% FS (conversion pressure <---> force)
-
LDM 70 - Digital Test Pressure
Gauge, accuracy 0.25% F.S. or 0.5% F.S.
- LPP-MANO - analogue precision pressure gauges DS 100
(4"), several ranges, accuracy 1% f.s.
We offer complete pressure calibration systems "
LPP-Kit"
with
- Calibration Test Pump
- Reference Pressure Gauge / Pressure Calibrator
- Test hose
- Set of adapters
- Carrying case

|
|