Portable Dry Block Temperature
Calibrator Type
JOKER+ |
 |
|

Temperature Dry Block Calibrator
JOKER+
- Range:
ambient temperature
up to +350°C
- for mobile and
stationary use
- micro processor operated
- weight: only 4.1 kg
- including
water proof
hard case
|
 |
Datasheet
Operating Manual
Typical Applications:
- Test and Calibration of all kind of thermometer.
- Test and Calibration of thermostats.
- Optional with RS232-interface: computer-aided calibration
|
Technical Data: |
Temperature range:
(at 20°C ambient temperature) |
+20°C...+350°C |
| Temperature stability: |
±0.1°C (at 300°C) |
Temperature uniformity:
(at 200°C) |
radial:
±0.35°C
(at 40 mm distance from bottom of holes)
axial: ±0.6°C
(at 60 mm distance from bottom of holes) |
| Stability indicator: |
no |
|
Display accuracy: |
±0.4°C |
| Display resolution: |
0.1°C |
| External
probe reading: |
no |
| Temperature units: |
°C |
|
Interface: |
optional
RS232 |
| Mean heating time: |
15°C per minute |
| Mean cooling
time: |
10°C per minute |
| Slope (ramp function): |
no |
|
Switch test: |
yes |
| Insert hole depth: |
130 mm |
|
Well diameter: |
26 mm |
| Dimension: |
215 x 120 x 255 mm |
| Weight: |
4.1 kg |
| Power supply (50/60 Hz): |
115 or 230 VAC |
|
Power consumption: |
300 VA |
Weight including hard case and accessories: 8.1 kg NOTE: the stated accuracy is valid for ambient temperature
+10°C...+40°C and is guaranteed for 1 year. Scope of standard
Supply:
- Dry Block Temperature Calibrator
JOKER+
- Electr. power cable
- Fuse kit
- Electr. connection cables for thermostat tests
- Operating manual
- Certificate of calibration, traceable
- Tweezers to change insert
- Insert with 3 holes (4.5 - 6.5 - 9.5 mm diam.)
- Insert with 2 holes (6.5 - 12.5 mm diam.)
- Insert without hole (to be drilled by customer)
- Hard case, water-proof, with foams

Optional Accessories:
- Insert with holes as per customers request
- RS232 interface
- RS232 cable
- Reference thermometer Pt 100 with DKD certification, as external reference
|
|