Portable Temperature Calibration Bath
FLUID 100
FLUID 200
FLUID 200-H |
 |
- for mobile and stationary use

- micro processor controlled (PID)
- ramp function
- thermostat test
- type
FLUID...-2I: with 2 inputs for external reference and test
specimen
- Recommended for the calibration of glass thermometer
- OPTIONAL: conversion kit to Dry Block Temperature Calibrator
Typical applications:
- Calibration and checking of all kind of thermometer
- Calibration and checking of thermostats
- computer aided calibration via RS232 interface
Technical Data:
| LR-Cal |
FLUID 100 |
FLUID 200 |
FLUID 200-H |
Temperature range:
(at 20°C ambient temperature) |
-12...+125°C |
+20...+200°C |
+20... +220°C |
|
Temperature stability: |
±0.02°C, at -5°C |
±0.02°C, at 150°C |
| Uniformity
Radial: |
±0.04°C, at 0°C |
±0.05°C, at 150°C |
| Uniformity
Axial: |
±0.05°C, at 0°C |
±0.05°C, at 150°C |
| Stability
indicator: |
audio+visual |
| Display
accuracy: |
±0.15°C |
| Display
resolution: |
0.1°C / 0.01°C |
| External
probe reading: |
Optional (version ...-2I) |
| Units of
measurement: |
°C / °F / K, switchable |
| Interface: |
RS232 |
| Mean
heating time: |
12°C per minute |
10°C per minute |
| Mean
cooling time: |
5°C per minute |
| Slope
(ramp function): |
yes |
| Switch
test: |
yes |
| Fluid: |
Silicone oil 200C5 |
<80°C: water
<125°C: silicone 200C5
<220°C: silicone 47V100 |
| Bath reservoir diameter: |
54 mm |
| Bath reservoir depth: |
150 mm |
| Power supply
(50/60 Hz): |
230 VAC (optional: 115 VAC) |
| Power consumption: |
300 VA |
500 VA |
| Fluid bath material: |
aluminum |
| Fluid bath volume (appr.) |
400 cm³ |
| Homogeny temperature via: |
magnet mixer, adjustable |
| Case material: |
metal |
| Dimension: |
160 x 340 x 330 mm |
| Weight: |
8 kg net |
NOTE: the stated accuracy is valid at ambient temperature +23°C and is
guaranteed for 1 year.
Scope of standard Supply:
- Portable fluid bath
FLUID
- Electr. power cable
- Fuses kit
- Electr. connection cables for thermostat tests
- Operating manual
- Certificate of calibration, traceable
- Bottle 500 cm³ with silicone oil
- Support for glass thermometer
- Closing top used for transport
- Soft bag
Additional for type
FLUID ...-2I:
- Version with data acquisition card and two input devices to measure
Pt 100 (3-/4-wire), Range -100...+700°C, Accuracy ±0.3°C
Thermocouple type J, Range 0...1000°C, Accuracy ±1°C
Thermocouple types K/N/R/S, Range 0...1300°C, Accuracy ±1°C
- Set of black and red connection cables with plugs
Optional Accessories:
- External reference probe, Pt 100 sensor, optional with DKD certification
- PC-Windows Software
AQ2sp and RS232 cable
With the software
AQ2sp it is possible to remote-control the
calibrator with a PC. Automatic or manual calibration of one or more test
specimen, stress tests, life cycle tests, thermostat tests. Printing of
calibration certificates.
- Extension tube to increase the dept of the liquid tank of
FLUID 100 and
FLUID 100-2I:

Usable diameter 60 mm, usable dept 230 mm
Minimum temperature: -9°C at 20°C ambient temperature
If used with silicone oil 200C5: Temperature range -9...+130°C at ambient
temperature +20°C.
Horizontal temperature uniformity: ±0.10°C (measured at 50 mm from the
bottom)
Vertical temperature uniformity: ±0.10°C (±0.15°C at temperatures below
0°C), over the dept of 150 mm from the bottom
- Extension tube to increase the dept of the liquid tank of
FLUID 200,
FLUID 200-2I,
FLUID 200-H
and
FLUID 200-H-2I:

Usable diameter 60 mm, usable dept 230 mm
If used with silicone oil 47V20: Temperature range 90...200°C at ambient
temperature +20°C
Horizontal temperature uniformity: ±0.20°C (measured at 50 mm from the
bottom)
Vertical temperature uniformity: ±0.10°C, over the dept of 150 mm from the
bottom
- Cooling coil in stainless steel for
FLUID 200,
FLUID 200-2I,
FLUID 200-H and
FLUID 200-H-2I:

The cooling coil shortens the cooling time, also the minimum calibration
temperature can be decreased.
- Conversion kit for fluid bath
FLUID (all models)
to get a Dry
Block Temperature Calibrator

Temperature range for models
FLUID 100 and
FLUID 100-2I:
-10...+125°C
Temperature range for models
FLUID 200 and
FLUID 200-2I:
ambient temperature
to +160°C
Block in aluminum, diameter 60 mm, usable dept 170 mm
Heating time from -10°C to +110°C: 45 minutes
Cooling time from +20°C to -10°C: 47 minutes
Stability: ±0.04°C
Vertical temperature uniformity at 0°C: ±0.03°C; at 80°C: ±0.06°C
Block with 9 holes 4.0 - 4.0 - 4.5 - 5.5 - 6.5 - 6.5 - 8.5 - 10.5 - 12.5 mm
or block without holes (to be drilled by customer).
|
|