Differential Pressure Deadweight Tester
CPB5000-DP
Ranges 0.03 bar to 1000 bar
2.9 to 15000 psi |
 |

Differential Pressure Deadweight Tester
CPB5000-DP
Differential Pressure Balance
Primary Standard for Differential Pressure

Datasheet

Operating Manual
- Ranges from 0.03 bar to 100 bar:
pneumatic operated
0.435 psi to 1500 psi
- Ranges from 0.2 bar to 1000 bar:
hydraulic operated
2.9 psi to 15000 psi
- Accuracy (total uncertainty of measurement) up to
±0.008% of measured value
- Tester with two piston-/cylinder-units and two weight
sets
- One spindle pump for generating the static line
pressure, another one for generating the differential pressure
- Options: DKD Certificate of Calibration
Deadweight Tester (Pressure Balances)
are the most accurate instruments for the calibration of electronic or
mechanical pressure measuring instruments. The direct measurement of
pressure, according to its definition as a quotient of force and area, and
the use of high-quality materials result in small uncertainties of
measurement and an excellent long-term stability of five years. For these
reasons, deadweight testers have already been used in calibration
laboratories of industry, national institutes and research labs for many
years.
The differential pressure deadweight tester
CPB5000-DP allows a real calibration of differential pressures,
based on static line pressure.
The
CPB5000-DP includes two common pressure
balances of the basic series
CPB5000. For the
calibration of differential gauges at a static line pressure, the system can
be separated (via bypass-valve).
One system is used to generate the line pressure. The second one supplies
the differential pressure (including line pressure).
For the basic specification, see type
CPB5000.
Optional Accessories
- Set of trim-masses for generating intermediate values
(weights 1 mg to 50 g)
- Oxytester (Purifier) for measuring instruments where
the filling medium should not come into contact with the deadweight
tester
- Dirt trap for test objects that are very dirty in order
to avoid the penetration of dirt particles into the deadweight tester
- Set of adapters in a tool case with 5 most commonly
used threaded adapters and a spanner (flats 32 and flats 14)
- Cleaning Kit "CPB5000-PN-RS" for pneumatic pistons
- Cleaning Kit "CPB5000-HY-RS" for hydraulic pistons
|
|
|
 |
|
|
|
Set of Adapters |
 |
|
|
|
Oxytester & Dirt trap |
|
Included in Standard Supply
- Basement with robust protective cover
- One inlet pressure pump and two spindle pumps
- Clamp for specimens
- 2 piston/cylinder-systems
- 2 sets of stainless steel weights in carrying case
Technical Data:
Measuring
Range bar |
Measuring
Range psi |
Version |
-0.03...-1
(Vacuum) |
-0.435...-14
(Vacuum) |
Pneumatic
(suitable for clean dry air or nitrogen - other media on
request) |
|
0.03...2 |
0.435...30 |
|
0.2...10 |
2.9...150 |
|
0.4...50 |
5.8...500 |
|
0.4...100 |
5.8...1000 |
| |
5.8...1500 |
| 0.2...60 |
|
Hydraulic
(as a standard for special oil - 1 litre included in
scope of delivery - other media on request) |
|
0.2...100 |
2.9...1000 |
| 1...250 |
14.5...5000 |
|
1...400 |
29...10000 |
| 2...600 |
29...15000 |
|
2...100 |
|
Accuracy (uncertainty):
±0.015% of measured value (%RD). The accuracy is characterized by the deviation
span, which is the sum of the systematic error and the uncertainties of
measurement. Long-term stability is not taken into account. Please note that
corrections might be required if the instrument is used without the optional
Intelligent Calibration Module - see operating manual.
As an option, systems with increased accuracies up to 0.008% of reading are
available
| Recommended accessory:
Hand Suction Pump HAP 02
Datasheet
for suction of fluids, e.g. oil, water, break fluid, etc.
Perfect for the suction of the operation fluid of deadweight
tester (e.g. before transport etc.). |
|
Other Deadweight Tester models:
-
CPB5000-HP:
High Pressure
Deadweight tester, up to 4000 bar / 60000 psi
-
CPB5000:
Deadweight Tester /
Pressure Balances
-
LDW 1000:
Deadweight Tester "Low Cost", Accuracy ±0.025%
|
|